Industrial resistivity meter HE-480R(W)

Overview

The HE-480R industrial resistivity meter features the high-precision temperature compensation function.
Its high-performance temperature compensation makes it ideal for sequential monitoring of ultra-pure water used in manufacturing processes in the semiconductor field, and many other fields, including the areas of electrical goods, foods and medicines. The measurement range is from 0 to 100.0 MΩ・cm.

Features

  • High-precision temperature compensation
    The temperature compensation element used is a platinum resistance thermometer (Pt 1000 Ω).
  • Freely adjustable standard temperature
    You can set the standard temperature for temperature compensation to a value of your choice, allowing it to meet the conditions you require, for processes such as measuring warm deionized water.
  • We make use of a full range of correction equipment
    Sensors are inspected prior to dispatch using correction equipment with ensured traceability. We have also put in place an after-sales support service, including on-site correction, and performing validation, for example by issuing correction certificates.
  • Uses lead-free solder
    To reduce the burden on the environment, lead-free solder has been used to mount the components on the printed circuit board.

Manufactured by HORIBA Advanced Techno

Specifications

Product name

Industrial resistance meter

Model

HE-480R

Measurement method

Electrode type (2-electrode method)

Cell constant

0.01 / cm

Temperature sensor
specifications

Platinum resistance 1000 Ω / 0 °C

Measuring range

Measurement unit

kΩ·m

0 to 2.00

0 to 20.0

0 to 200.0

0 to 1000 *

MΩ·cm

0 to 0.200

0 to 2.00

0 to 20.00

0 to 100.0 *

*: Measurable without temperature compensation

Temperature: 0 °C to 100 °C (Select your desired decimal point from 0, 1, and 2 digits.)

Repeatability Within

± 0.5 % of the full scale (in equivalent input)

Linearity Within

± 0.5 % of the full scale (in equivalent input)

Transmission
output

4 mA to 20 mA DC: input / output isolated type
Maximum load resistance: 900 Ω
Transmission output range: Freely selectable within the measurement range

Contact output

Outputs: 2 points
Alarm contact output (R1 and R2)
Contact type: relay contact, SPDT (1c)
Contact rating: 240 V AC, 3 A and 30 V DC, 3 A (resistance load)
Contact function: selectable from upper / lower limit operation (ON / OFF control), alarm, and maintenance.

Transmission
output hold feature

  • Previous value hold
  • Optional value hold
  • Continuous

Selectable from the above (However, only the previous value hold is available in the maintenance Mode.)

Self-diagnosis
function

  • Sensor diagnosis
  • Out of the measurement range
  • A/D converter scale over
  • Converter error

Temperature
compensation

  • Based on the temperature characteristics of extra deionized water (reference temperature: 25 °C)
  • Based on the reference temperature and user-defined temperature coefficient (reference temperature: 5 °C to 95 °C)
  • No temperature compensation

Temperature - compensated range

0 °C to 100 °C

Power supply

Rated voltage 100 V to 240 V AC, 50 / 60 Hz, 10 VA (max.)

Structure

Indoor-use panel installation type
Panel: IP65 dust and water proof structure

Protective structure

Panel: IP65 (IEC60529, JIS C0920)

Conforming standards

CE Marking
FCC Rule

External
dimensions

48 (W) mm x 96 (H) mm x 115 (D) mm
Case depth: approx. 105 mm (when panel-mounted)

Mass

Approx. 400 g

Compatible
sensors

ERF-series, cell constant 0.01 / cm

Schematics

Dimensions of HE-480R(W)

Related Products

Carbon Sensor Resistivity Meter HE-960R-GC(W)

Carbon sensor resistivity meter. Measurement range: 0 - 100.0 MΩ・cm

2-channel Carbon Sensor Resistivity Meter HE-960RW-GC(W)

Resistivity meter having the 2-channel carbon sensor. Measurement range: 0 - 100.0 MΩ・cm