Accurate pH measurement
There are various precautions to perform more accurate pH measurement. To perform more accurate and precise measurement, refer to the following descriptions.
The pH measurement and the solution temperature
The pH value of a solution has the inherent temperature characteristics. For the pH measurement, the temperature shall be controlled as accurate as possible.
*The measurement with the constant temperature bath and temperature compensation function*
Use the constant temperature bath to perform the calibration/measurement steps at the same temperature of the solution.
Use the temperature compensation function of the meter during the calibration/measurement steps.
However, if the manual temperature compensation function is equipped, the input operation to the meter is available.
*The measurement with the constant temperature bath*
After the calibration in the constant temperature bath, keep securely the temperature of the bath and the sample at 25℃, and then perform the measurement. In this case, the temperature compensation and the temperature conversion are not necessary
*The measurement at the ambient temperature*
(1) During the calibration/measurement steps, use the temperature compensation function.
(2) If the manual temperature compensation function is equipped, the input operation to the meter is available. In addition, during the calibration/measurement steps, the solution shall be controlled at the same temperature as much as possible.
Precautions before immersing the electrode in the sample
(1) Open the replenishment inlet for the internal solution of the reference electrode.
(2) Fill the reference part sufficiently with the internal solution.
(3) When bubbles are in the internal solution, remove them by shaking the electrode.
(4) Use the thoroughly rinsed electrode.
Actions for more precise measurement
(1) Keep the constant temperature of the calibration solution and sample solution in the constant temperature bath (in order to reduce the error caused by the temperature change).
(2) Perform the measurement in the sealed box (in order to avoid the outside-air influence).
(3) Use an airtight measuring container (in order to avoid the outside-air influence).
(4) Rinse the measuring container with pure water, and then rinse it with the sample again.
(5) Stir the sample slowly during the measurement.
(6) When measuring with the discrete electrode, place the reference electrode higher than the glass electrode.
(7) When the buffer capacity is particularly low, perform the measurement with a flow cell system.
(8) Conditioning the electrode in advance (in order to let the electrode get to know the sample).