Application
FACT 220
Overview
FACT 220 is a testing device for automatic verification of the calibration of the sensors installed in the FQ-2100DP system. The measuring unit of FACT 220 consists of a density and fuel flow sensor that allows to check the entire measuring range from 0.2…220l/h. The volumetric flow rate is set by a remote controlled valve and can be set automatically with predefined points for a fast verification and high precise quality of the measurement result. The FACT 220 is operated via a control panel; hereby you can start the validation procedure and status as well as statistical values can be traced with.
There are two possibilities to connect the FACT 220, to the inlet circuit of the FQ-2100DP or like a specimen in the outlet circuit. The flow deviations between FQ-2100DP system and reference will be valuated with a status message. At the end of the testing procedure a test report will be created. An additional tank is filled with preservative for post-treatment after measurement. The unit will be connected over Ethernet-Interface with FQ-2100DP for data transfer and remote control of all functions like manual and automatic testing and purging. The system is designed for commercially available fuels, Ethanol up to 100%, Kerosine and Biodiesel.
Features
- High exactness of the measurement in the whole measurement range 0.2…220l/h
- User-friendly handling through Touch-Panel; test results are displayed
- Quick check of individual measurement points through manual test mode
- Summation counter for fuel flow volume and hours of operating service
- Different connection possibilities for checking of the whole system or flow sensor by itself
- Test report with statistical values
- Online evaluation of actual measurement uncertainty through continuous calculation of statistic values ( moving average, standard deviation)
- Predefined measurement values for different sensors
Manufactured by HORIBA
Specifications
Measurement range volume flow | 0.2…220l/h |
Measurement range density | 0.5…2g/cm³ |
Fuel pressure of system | 2…2.5 bar |
Supply voltage | 230V |
Power frequency | 50/60Hz |
Nominal current | 2A |
Size | 675mmx1120mmx540mm |
Weight | 90 kg |
Gradient Ambient Temperature | <0.02 [K/min] |
Connection Power | 0,5 kVA |
Measurement uncertainty Calibration factors Volume flow range 0,2…5l/h range 5…80l/h range 80…220l/h |
±0,105% ±0,075% ±0,05% |
Measurement uncertainty density | ±0,02% |