MEXA-700 Lambda
Overview
With its wide measurement range and rapid response time, MEXA-700 Lambda is ideal for monitoring dynamic changes in all engine types, including lean-burn. Applications range from R&D to engine tuning.
At one-third the size of previous models, MEXA-700 Lambda can be:
- Mounted on the dashboard of a vehicle for on-road measurements.
- Stacked one on top of another for multi-channel measurements.
- Mounted in a 19-inch rack cabinet with other instrumentation.
After an initial 1-minute warm-up time, a single key-press starts the measurements. Readings are displayed by an LED on the front of the analyzer. An RS-232 interface and 0-5 V analog output allow data collection by a host computer.
Features
Applications
- * Lean-burn and/or diesel engine testing
- * Engine research and development matching
- * Engine tuning
- * Catalyst development
- * Exhaust gas adjustment
- * On-board evaluation
- * ECU adjustment
Features
- Compact - can be mounted on a dashboard, on a table-top, or in a 19-inch rack.
- Quick response - typical 80 ms.
- Wide range measurement - suitable for any engine type, including lean-burn.
- Accuracy of UEGO sensor guaranteed. Every unit undergoes thorough inspection.
- Stackable for multi-channel measurement.
Options
- AC power adapter: Input 100-240 V AC(50/60 Hz), output 13 V DC
- Rack mounting kit: For one or two MEXA-700 Lambda units, with an AC power adapter
- Additional cable for sensor: 5/20 m
- Calibration unit: For generating the calibration gas
Manufactured by HORIBA
Specifications
MEXA-700 Lambda Specifications | |
Ranges | Air/Fuel ratio: 0.0-99.9 (at H/C=1.85, O/C=0.00) Excess air ratio (Lambda): 0.00-9.99 (at H/C=1.85, O/C=0.00)
Oxygen (O2): 0-25 vol%
|
Response Time | T63 within 80 ms typical; 150 ms maximum |
Accuracy> | Model gas of room temperature, atmospheric pressure is used, in from 9.5 to 60 A/F range
Within ± 0.3 A/F (at approx. 9.5 A/F) |
Warm-up> | Approx. 1 minute after start-up of the analyzer |
Acceptable> | For sensor: |
Calibration> | For lean condition: O2 or atmospheric air |
Sample Gas Conditions | Without water drop and mist |
Ambient Conditions | For main unit: 5 to 45° C; less than 80% R.H. |
Dimensions> | W x D x H (excluding protrusions) 130 x 170 x 75 mm Approx. 1.6 kg / 3.5 lb |
Power | 10-15 V DC, 50 W |
Outputs | Analog: 0-5V DC, 1 channel, full scale is changeable |
Accessories | Included |