ABX Micros CRP 200
Resumo
Simultaneous CBC & CRP
- Single tube operation : Whole blood (EDTA)
- Single microsample : CBC – 10uL; CBC + CRP - 18 µL
- Fast and efficient process : No centrifugation
- Inflammation testing : CBC and CRP
- Compact analyzer
- Integrated reagents
- Reliable results
- "Zero maintenance" compact data management systemmpedance technology, photometry, immunoturbidimetry
Caracteristicas
ABX Micros CRP 200
- WBC, RBC and PLT : Impedance
HGB : Spectrophotometry (Cyanmethemoglobin method)
HCT : Numeric integration
CRP : Latex immuno turbidity rate method
MCV, MCH, MCHC, RDW, MPV, PCT* and PDW*.
Calculations.
*: RUO parameters (Research Use Only)
Manufactured by HORIBA
Especificações
Technical
Power requirement | Power supply from 100 Vac to 240 Vac +/- 10%
|
Dimensions and Weight | Weight : 40 lbs. |
Memory | QC (CBC) and QC (CRP) modes: 30 samples |
Samples
Parameters | CBC mode 18 parameters with graphics for RBC, PLT and WBC populations. CBC & CRP mode 18 parameters & CRP |
Sample-Volume | CBC mode : 10 µl |
Measurement item | WBC, RBC, HGB, HCT, PLT, CRP (C-reactive protein) MCV, MCH, |
Sample processing | On open tube with 4 position tube holder (QC, 500 µL, 3 mL, 5 mL) |
Throughput analyses | CBC mode: Within 1 min, 15-sec (48 samples/hour) |
*: RUO parameters (Research Use Only)
Results
Linearity
Parameters | Linearity |
|---|---|
WBC | 0-80 x 10³/mm3 |
RBC | 0-7,5 x 106/mm3 |
HGB | 0-23 g/dl |
HCT | 0-62.4% |
PLT | 0-900 x 10³/mm3 |
CRP (Whole Blood) | 2-200 mg/L |
CRP (Serum) | 2-150 mg/L |
Precision
Parameters | Presicion (cv %) | Range |
|---|---|---|
WBC | < 2,5% | 4.75 - 12.3 x 103/mm3 |
RBC | < 2% | 4.0 - 5.0 x 106/mm3 |
HGB | < 1.7% | 12.0 - 16.6 g/dl |
HCT | < 2% | 37 - 50% |
PLT | < 5% | 150 - 355 x 103/mm3 |
CRP | < 10% | 0-10 mg/L |
< 5% | > 10 mg/L |