INM-700 In-Situ NOx/O2 Monitor
Übersicht
HORIBA's INM-700 in-situ gas analyzer is designed to meet the requirements of coal-fired power generation plants for cost-effective, low maintenance NOX monitoring. The analyzer uses a ceramic diffusion filter and Zirconium Oxide sensor directly in the gas stream, which eliminates the need for an extraction process for sample collection, and results in lower maintenance. The INM-700 is ideal for inlet SCR control, NOX measurements, and burner balancing/combustion tuning in boiler plant operations. The INM-700 has the advantage of lower installation, maintenance, and capital costs, and a faster response time. In addition, for compliance strategies, the ceramic filter allows measurements to be taken near the combustion point of the furnace. This technique utilizes combustion modification techniques such as:
- Low NOX burners (LNB)
- Over-fire air injection (OFA)
- Gas re-burn systems
- Boiler tuning
- Combustion optimization programs
Applications
- SCR control
- Inlet/outlet NOX measurement
- Replacement of FGAS shelters
- Burner balancing/combustion tuning
- Improved burner efficiencies and reduction of NOX
- Identifying problem areas in the furnace due to gas stratification linked to specific burner columns.
Merkmale
- A NOX sensor using a new technology for coal fired applications which can improve control of the ammonia injection rate in SCR systems.
- A practical NOX sensor with a fast response time that can be directly installed in the combustion gas stream.
- Extremely robust sensors that are long lived and capable of withstanding elevated temperatures.
Hergestellt von HORIBA
Spezifikationen
table>
Model
INM-700
Principle
Zirconia type
Measurement range
Nox: 0 to 100 ppm(minimum),
0 to 1000ppm(maximum)
O2: 0 to 25%
Repeatability
Nox: ±1.0% of F.S.
O2: ±1.0% of F.S.
Drift
±2.0% of F.S./week
Response time(T90)
10 sec. or less
Display
LED digital display (effective 4 digits) for measuring value
LCD display for parameter setting
Output signal
Nox: DC 4 to 20 mA(maximum load 600 )
O2: DC 4 to 20 mA(maximum load 600)
Corrected Nox: DC 4 to 20mA
(allowable load resistance: 600 )
Calibration
Probe calibration (Automatic or manual)
Warm up time
approx. 5 min
Sensor cable
Max. 50m (164ft.)*1
Location
Indoor or outdoor
Operational temperature
- 10 to 50°C(23 to 122°F)
Humidity
90% RH(no moisture condensation)
Power
AC100V to 240V (50/60Hz)
Power consumption
approx. 60VA
Mass
Probe unit: approx. 10 kg (approx. 22lbs)
Controller unit: approx. 20kg (approx. 44lbs)
Cooling air consumption:
Stack gas application
(Instrument air)
400°C (752°F): approx. 70 L/min.(2.47 cfm)
350°C (662°F): approx. 50 L/min.(1.77 cfm)
300°C (572°F): approx. 40 L/min.(1.41 cfm)
200°C (392°F): approx. 40 L/min.(1.41 cfm)
Purge air
(Instrument air)
Pressure air: about 300kPa
Consumption air: 3 L/min.