Elementar Analysatoren
XGT-5700WR - Röntgenfluoreszenz-Gerät
Übersicht
XGT-5700WR are new lineup of XGT-5000WR series and equipped with new optical system to analyze hazardous elements with high sensitivity. XGT-5700WR support for WEEE/RoHS, ELV and Chinese RoHS
These models have 3mm XGT with multi-filter with 4 specific filters (Pb/Cd/Cr,Cl/5 WR elements) and 100 µm XGT as standard and 10 µm is optional instead of 100 µm. Secondary filter(between sample and detector) with ON/OFF positions are also equipped. Improved detection sensitivity is realized through the use of a dual filter. Compliant with WEEE/RoHS, ELV Chinese RoHS directives
XGT-5700WR is equipped with dewar of liquid nitrogen. XGT-5700WR Type S/SL model with big sample room are lineuped.
Merkmale
Increased detection sensitivity with a new optical system
3mm X-ray irradiation diameter and dual filter system (Primary and Secondary filters)- Capable of both minute area analysis and high sensitivity analysis
Minute area analysis: 10(with no filter) or 100 µm (with no filter) X-ray irradiation diameter
High sensitivity analysis: 3mm X-ray irradiation diameter and dual filter system
Maintains the same ease of operation as previous systems to promote efficient on-site work
Preset measuring conditions for materials ranging from resins to metals
A lineup of data management software is available that is compatible with various data format
Can be used to analyze hazardous elements in Ni and Sn plating (When using the multilayer film FPM option)Internal analysis using element mapping for minute area and transmission image
Applications ranging from point measurement to surface measurement
Hergestellt von HORIBA
Spezifikationen
Model | XGT-5700WR |
Principle | Energy dispersive X-ray fluorescence analyzer |
Element | Na to U (in sample atmosphere) |
Sample | Plastic, metal, paper, liquid such as paint and ink, biological samples etc |
Sample chamber atmosphere | Atmosphere |
XGT Element Spatial Resolution | 3mm+100µm or 3mm+10µm |
X-ray tube | 50kV/1mA, Rh target |
Vacuum probe | Standard |
Primary X-ray Filter | Ø3mm:Multifilter with 4 specific filters(Pb/Cd/Cr,Cl/5WR elements) |
Secondary X-ray filter | Standard |
Detector | High purity Si detector |
Stage | 100mm×100mm(Maximum mapping area) |
Sample chamber | 400×350×40mm(Additional special order is possible) |
Optical image | Entire image: 100mm×100mm (Standard:400,000 pixels, Optional: 2million pixels) |
Analysis functions | Qualitative: Automatic qualitative function, BG display, ROI classification by color, Mutching function |
Data management(Optional) | Excel® data management software, Ecological procuremet supporting software and an Inspection report output |
Power Supply | AC100, 120, 220, 240V 50/60Hz |
Power Comsumption | 1,3kVA or lower |
Weight | Approx 280kg |
Outer dimensions | Analyzing unit: 680(W)833(D)670(H)mm |