IT-450 Series Non-contact infrared thermometer

Vue d'ensemble

This non-contact infrared thermometers achieve miniaturization like that of an ultra-compact photoelectric switch, and focusing of the measurement field into a single spot. They enable measurement without damaging the object, whether the object is stationary, rotating or moving, and can flexibly handle a wide range of temperature control tasks in production lines, such as line stopping and heater control. Six models are available to suit the application.

Fonctions

  • 0.4 sec / 95%, high response speed
  • Measure on very little and long distance surface with high precision according to spot measurement on measured view.
  • To be realized wide measuring range.

Fabriqué par HORIBA

Caractéristiques

     

IT-450S

IT-450L

IT-450FH

IT-450NH

IT-450F

IT-450N

Measured temperature range

0 °C to 500 °C

0 °C to 1000 °C

-50 °C to 500 °C (setting cange in programer is required at less than 0 °C)

Analog output*1

4 mA to 20 mA / 0 °C to 500 °C Load impedance: 300 Ω or less

4 mA to 20 mA / 0 °C to 1000 °C Load impedance: 300 Ω or less

4 mA to 20 mA / 0°C to 500 °C Load impedance: 300 Ω or less

Accuracy*2

±(3+0.1% of output range)  °C or less (0 °C to 300 °C)

±(2% rdg-3 + 0.1% of output range) °C or less (300 °C to 500 °C)

±(2+0.1% of output range) °C or less (0 °C to 200 °C)

±(1% rdg + 0.1% of output range) °C or less (200 °C to 1000 °C)

±(10% rdg +2+0.1% of output range) °C or less (-50 °C to 0 °C)

±(2+0.1% of output range) °C or less (0 °C to 200 °C)

±(1% rdg + 0.1% of output range) °C or less (200 °C to 500 °C)

Repeatability

1 °C or less

0.5 °C or less (0 °C to 500 °C or less)

1 °C or less (500 °C to 1000 °C or less)

1 °C or less

0.5 °C or less

Temperature drift

±(0.2 °C in the unit of  °C + 0.01% of output range in the unit of  °C) or less

±(0.25°C in the unit of °C) or less (0°C to 500°C or less)

±(0.5 °C in the unit of  °C) or less (500 °C to 1000 °C or less)

±(0.5°C in the unit of °C) or less (-50°C to less than 0°C)

±(0.25 °C in the unit of  °C) or less (0 °C to 500 °C or less)

Measurement field / distance

δ3/30

δ8/100

δ35/1000

δ40/500

δ35/1000

δ40/500

Mass

Approx. 120g

Approx. 150g

Approx. 120g

Approx. 150g

Approx. 120g

Emissivity

Set to 0.95 at factory

Spectral response

8 µm to 16 µm

Response time

0.4 seconds or less (95% response, Moving average data value = 1)

Ambient temperature/humidity

Temperature: 0 °C to 55 °C, relative humidity: 35% to 85% (No condensation)

Power supply

12V to 24V DC

Power consumption

40mA or less (45 mA or less when programer is connected.)

Cable length

2 m

Unit case material

NYLON 6

Water-resistance

IP X2

*1: Optional programmer enables freely output scale settings
*2: Ambient temperature: 23 °C ± 5 °C, Ambient humidity: 55% ± 20%,Emissivity: 1.00
rdg: Reading value