Corresponding detector | ZH-10 | ZH-40 | ZH-100 | ZH-500 |
---|---|---|---|---|
Measurement range | 0 - 10.00mg/L | 0 - 40.00mg/L | 0 - 100mg/L | 0 - 500mg/L |
Supply temperature | 5 - 30 deg.C | 5 - 80 deg.C | ||
Chemical contact material of detector | PFA, PTFE, quartz glass | |||
Supply pressure | Within 0.5 MPa | |||
Repeatability | +/- 0.2% of full scale | |||
Linearity | +/- 1.0% of full scale | |||
Transmission output | Number of output: 2 | |||
Contact output | Number of output: 5 | |||
Contact input | Number of input: 2 | |||
Communication output | RS-485 | |||
Power supply | Rated voltage 24 +/- 10% V DC, 15W |
Schematics