ABX Pentra 120 Retic SPS

概要

  • Microsampling of 50 µl whole blood from primary tubes.
  • User-definable smearing criteria.
  • Choice of staining protocols.
  • Validation station with cytology atlas.
  • Up to 120 samples/hour.

特長

ABX Pentra 120 Retic SPS Concept and technology

  • Methods of measurement: Cytochemistry, impedance, absorbance and flow cytometry (Eosinofix* and LMNE matrix).
  • Reticulocyte profile:Thiazole orange, fluorescence, impedance, Argon-ion laser (Retix* and retic matrix).
  • Balance Concept: Automatic control of the leukocyte count based on 3 independent channels.
  • DHSS*: Focused flow cytometry and sequential measurement (impedance and absorbance).
  • Automatic re-run of samples : Selective and programmable (hematology parameters and smears).
  • Rotative mixing of samples : Smooth and efficient
  • Wedge smear tape cartridge: Optimized slide preparation and no contamination.
  • Dedicated slide staining wells : No contamination and minimum dye consumption.
  • Liquid valves* ABX (no pinch valves): Reliability and volume accuracy

*ABX patents


製造会社: HORIBA ABX SAS

仕様

Technical

Weight

170 kg

Power supply

90/240 V - 50/60 Hz - 800 VA.

Dimensions

50 x 91 x 70 (H x W x D) cm (Pentra 120 module).
Pentra 120 + SPS = 160 cm.

Electrical safety standards

IEC 601-1 CE Mark.

Throughout

120 tests/hour

Physical

Temparature range of 18-32°C. 20-85% relative humidity (non-condensing).

Performance

Produces classic wedge smear with significant monolayer.
No adjustement, computerized smear optimization. Classic wedge smear 15 smearing profiles.

Specific parameters

Drop volume, smearing angle, smearing speed.

Staining Rotor

32 staining wells.

 

Samples

120 tests/hour (closed tube)

Processing

Up to 120 sliedes per hour.
Smearing with or without staining upon user defined criteria.
Rerun slide capacity (reflex testing).
"Stat" for staining pre-smeared slides.

Staining

May Grünwald-Giemsa.
Wright.
Wright-Giemsa.

Volume

Closed tube mode : 50 µl

Operating sample mode

Vertical rack loader.
End-over-end rotation for smooth sample mixing.
Maximum load: 15 racks (150 tubes).

Identification

Auto-printing of demographics (user definable) on slide with resilient ink.
Text capacity of 3 printed lines (user definable).

Parameters (36)

WBC, RBC, HGB, HCT, MCV, RDW, MCH, MCHC, PLT, MPV, PCT, PDW % and #. LYM, MON, NEU, EOS, BAS0. RET %, RET #, RETH %,RETM %, RETL %, IMM %, CRC%, IRF %, MRV, MFI%

Plus

ALY = Atypical Lymphocytes (% and #)
LIC= Large Immature Cells (% and #)

Graphics

High resolution leucocyte differential matrix
RBC, PLT, and BAS histograms.
RES and FLUO histograms.
Retic matrix.

Stability

Up to 48 hours

 

Results

Parameters

Presicion (cv %)

Linearity

WBC

< 2% (at 10.10³/mm³)

0-150.10³/mm³

RBC

< 2% (at 5.0.106/mm³ )

0-8,1.106/mm³

HGB

< 1% (at 15 g/dl)

0-25 g/dl

HCT

< 2% (at 45%)

0-80%

PLT (whole blood)

< 5% (at 300.10³/mm³)

0-2000.10³/mm³

PLT (PLT concentrate)

0-4500.10³/mm³

LYM

3.58% (at 23,3%)

0-100%

MON

8,11% (at 8,2%)

0-100%

NEU

2% (at 65,4%)

0-100%

EOS

11,45% (at 2,4%)

0-100%

BAS

30%

0-100%

Parameters

Presicion (cv %)

Linearity

WBC

< 95%

0-150 x 103 / mm3

RBC

< 95%

0-8,1.106/mm³

HGB

< 95%

0-25 g/dl

HCT

< 95%

0-80%

PLT (whole blood)

< 95%

0-2000.10³/mm³

PLT (PLT concentrate)

< 95%

0-4500.10³/mm³

LYM

< 95%

--

MON

< 95%

--

NEU

< 95%

--

EOS

< 95%

--

BAS

< 95%

--

価格は仕様により異なりますので、別途お見積もりいたします。「見積・資料・お問い合わせ」よりお申し込みください。

カスタマーサポートセンターでは、お電話でも製品の技術的なご質問や仕様のご相談を承ります。

堀場製作所
カスタマーサポートセンター

フリーダイヤル 0120-37-6045
受付時間:
9:00~12:00/13:00~17:00
月曜日~金曜日(祝日をのぞく)

関連製品

ABX Pentra 120

26 parameters with 5 Part DIFF