510 Series Gas Analyzers

Overview

The 510 series of general-purpose gas analyzers provide continuous monitoring of concentrations of the specific sample gas. The analyzers can be operated from controls on the front panel or by commands from a remote computer. Measurement results are displayed on the front panel and are available to remote data logging systems through an industry-standard interface.

The 510 series can be used for a wide variety of analyses and tests, such as industrial process control and composition analysis, and fixed-source emissions monitoring. Each model is configured using one of four analyzers: the VIA-510 infrared analyzer, the CLA-510 chemi-NOX analyzer, and the MPA-510 O2 analyzer.

A high level of sensitivity is achieved through the use of a dual-beam NDIR, Chemiluminescence, and Magneto-pneumatic analysis methods. The NDIR analyzers use an infrared absorption method which offers superior sensitivity, selectivity, and stability. With the Magnetopneumatic analyzer, high-sensitivity measurements can be realized free of zero drift due to the use of N2 as the operation gas. The Chemiluminescence analyzers use a dryer unit equipped with automatic reseneation, so ambient air can be used as the source of O3. It also provides 7 ranges and a range of 1:100 (0-20/50/100/1000/2000ppm), making it easy to handle a wide variety of applications. These highly accurate, high performance analyzers are suitable for process monitoring and control.

Features

  • Selectable outputs: 0-1 VDC, 0-10 VDC, and 4-20 mA
  • Digital outputs: range setting information.
  • Digital alarms: indicate malfunctions or calibration failure.
  • Digital inputs: Zero cal, Span cal, Data hold.
  • Range inputs: 4 inputs BCD code.

Manufactured by HORIBA

Specifications

Model

Gas

Min.

Max.

High-sensitivity range

VIA-510

Carbon Monoxide (CO)

0-100ppm

0-100%

0-50ppm

VIA-510

Carbon Dioxide (CO2)

0-100ppm

0-100%

0-50ppm

VIA-510

Sulfur Dioxide (SO2)

0-100ppm

0-100%

0-50ppm

VIA-510

Methane (CH4)

0-100ppm

0-100%

0-50ppm

VIA-510

Nitrous Oxide (N2O)

0-100ppm

0-100%

0-50ppm

MPA-510

Oxygen (O2)

0-5%

0-100%

0-1%

CLA-510

Nitric Oxide (NO)

0-100ppm

0-5000ppm

0-20ppm

 

 Common specification in 510 series

 

Configuration

Desk top or 19 " panel mount

Ambient conditions

Temperature: 0 to 40 °C

Humidity: 90% RH or less

Power

100 / 115 / 220 V DC, 0 - 10 V DC, 0 - 16 mA or 4 - 20mA DC, isolated

Output signal

0 - 1 V DC, 0 - 10 V DC, 0 - 16 or 4 - 20mA DC, isolated

Display

LED digital display (effective 4 digits.)

Dimensions

430 (W) x 132 (H) x 550 (D) mm

16.9 (W) x 5.2 (H) x 21.7 (D) in

Sample gas requirements

Dust: No

Mist: Completely removed

Pressure: More than 50 mm H2O

Temperature: Ambient temperature

Moisture: 5°C saturation or below

Flow rate: 500mL / min

Exhaust gas: Discharged at ambient pressure

 

 

VIA-510

 

Measureing method

NDIR

Repeatability

±0.5% F.S. (±1.0% F.S. for high-sensitivity range)

Drift

±1.0% F.S. per day (±2.0% F.S. per day for high-sensitivity range)

Response time

10 sec. (electrical)

Operation gas

Not needed

Calibration gas

Zero: N2 or Air, Span: Span gas

Mass.

15kg, 33 lb

 

 

 MPA-510

 

Measureing method

Magnetopneumatic

Repeatability

±0.5% F.S. (±1.0% F.S. for high-sensitivity range)

Drift

±1.0% F.S. per day (±2.0% F.S. per day for high-sensitivity range)

Response time

10 sec. (electrical)

Operation gas

N2, Flow rate: 6 mL/min.

Calibration gas

Zero: N2, Span: Span gas or Air

Mass.

15kg, 33 lb

 

 

CLA-510SS

 

Measureing method

Chemiluminescence

Repeatability

±0.5% F.S. (±1.0% F.S. for high-sensitivity range)

Drift

±1.0% F.S. per day (±2.0% F.S. per day for high-sensitivity range)

Response time

10 sec. (electrical)

Operation gas

Not needed (ambient air used)

Calibration gas

Zero: N2 or Air, Span: Span gas

Mass.

20kg, 44 lb

 

 