APSA-360A/CU-1 Ambient Hydrogen Sulfide Monitor
Overview
APSA-360A/CU-1 Ambient H2S monitor measure SO2 converted H2S through oxidation catalyst based on Ultra Violet Fluorescence. In order to reduce measurement error by water concentration fluctuation in ambient, humidifier is set at sample line. It enables long-term stability according to stabilizing catalytic reaction.
Features
- Monitor H2S with high sensitivity range in ambient air continuously.
- Set humidifier at sample line to stabilize catalytic reaction.
Manufactured by HORIBA
Specifications
Measurement component | Hydrogen sulfide(H2S)in ambient air |
Measurement principle | Oxidation catalyst + Ultra Violet Fluorescence |
Measurement range | 0~0.1/0.2/0.5/1.0ppm |
Lower detectable limit | 4ppb |
Repeatability | ±3.0% of F.S. |
Linearity | ±2.0% of F.S. |
Zero drift | ±2% of F.S./day |
Span drift | ±5% of F.S./day |
Response time(T90) | 180 sec. or less (from CU-1 converter unit) |
Sample gas flow rate | approx. 0.8L/min |
Operation gas | None |
Indication | Measuring value, alarm |
Alarm | CU-1: Temperature error in catalyzer temperature error in converter, etc. |
Input / Output | 0~1V / 0~10V / 4~20mA (2 systems: either (1) momentary value and integrated of (2) moving average value) |
Temperature/Humidity | 0~40℃ 85% or less |
Power | AC100V±10V 50/60Hz |
Power consumption | CU-1: 150VA |
Dimension | CU-1: 430(W) X 550(D) X 221(H) mm |
Mass | CU-1: approx. 10kg |