MPA-31 Series Magnetopneumatic Oxygen Analyzer

Overview

The MPA-31 Series is a lineup of units that can control boiler combustion in a variety of industrial settings and monitor atmospheric oxygen concentrations. Using a magnetic pressure method that utilizes oxygen's high level of ordinary magnetism, the units are highly accurate and offer superior response time thanks to the use of an independent condenser microphone in the output device. In addition to being resistant to interferents, the units offer highly stable data output over the long-term for process control as they are not subject to temperature- and longitudinal time-induced zero drift. These explosion-resistant devices are suitable for a wide range of applications, from boiler combustion control and the monitoring of atmospheric conditions inside a variety of furnaces to controlling nitrogen purity at air separation plants and conducting quality management of municipal gas.

Features

The units employ a magnetopneumatic method that utilizes oxygen's high level of ordinary magnetism and realize high-speed response through the use of a condenser microphone output device and small cell capacity.The units are deliver continually high levels of measurement accuracy and have superior selectability with respect to oxygen due to proportioning of the magnetic susceptibility of coexistent gases.The units are theoretically not subject to zero drift thanks to the use of small amounts of N2 as a auxiliary gas (reference gas). The units are also capable of creating a deviation range when using 100% O2 and 21% O2.With two models-the d2G4 pressure explosion resistant MPA-31 and the fid3aG4 internal pressure explosion resistant MPA-31F-the series provides users with a wide range of gas measurement capabilities.


Manufactured by HORIBA

Specifications

Principle

Magnetopneumatic method

Range

0-1 to 0-100 %,max.3 ranges 

Range ratio

1:2:5,1:2.5:5,1:2:10,1:2.5:10,1:5:25,

Devation range(option)

90-100,95-100 or 16-21%

Enclosure

MPA-31:d2G4*
MPA-31F:fd3aG4*
*Approved by The Research Institute of Industrial Safety.

Repeatability

±1.0% of full scale

Zero/Span Drift

Zero
±1.0% of full scale*
Span
±2.0% of full scale week
*Guaranteed at normal ambient temp. ±5°C

Linearity

Within ±1.0% of full scale

Response time

(Standard)
90% response whithin 2 sec.
(under 5% renge or deviation range)90% response within 10 sec.

Ambient conditions

Temperature

0-40°C

Humidity

90% or less

Power

100/115 V AC,50 of 60±0.5Hz
(Consumption)
MPA-31:Approx.70VA
MPA-31F:Approx.120VA

Display

31/2-dight,LED digital display

Sample conditions(at the sample inlet)

Flow rate

1.5±0.5L/min

Pressure

15-25Kpa

Temperature

20-40°C

Refarence gas conditions(N2 standard)

Flow rate

10 mL/min

Pressure

49-98Kpa

Output

liner 0-16 or 4-20 mA DC,isolated,load resistance max.750Ω

Connection

1/Rc female for inlet and outlet

Materials in contact with sample

304 stainless steel,viton,gold,epoxy resin

Mass

Approx.60kg

Option

AIC Series Automatic Calibration System to be installed in non-hazardous area.