ESH-01-C-SN Sanitary Conductivity Sensor (Ferrule Clamp type) for Low Concentrations ESH-01-C-SN
Overview
The ESH-01-C-SN conductivity sensor, which is used in conjunction with the HE-960CW conductivity converter, is ideal for controlling conductivity in the purified water, pasteurized water, and water for injection that are used in the production of pharmaceuticals.
Features
- Tough structure that can withstand steam sterilization
Complies with the steam sterilization condition at 140°C / 0.6 MPa (60 minutes) - Electro-polished and mechanically polished, which satisfies the requirements for sanitary pipes.
- SUS316L, PEEK, and FKM are used as materials for contact with fluids
Manufactured by HORIBA Advanced Techno
Specifications
Model | ESH-01-C-S-SN 1.5 S |
|---|---|
Measurement principle | Alternative two-pole method |
Cell constant | Approx. 0.1 /cm |
Measurement range | 0 to 200 µS/cm (complies with the specifications of the conductivity converter) |
Measurement temperature | 0 to 100°C (complies with the specifications of the conductivity converter) |
Pressure range | 0 to 1 MPa |
Ambient operating temperature / humidity | 0 to 50°C, 95% RH or lower |
Structure | IP67 or equivalent |
Materials for contact with fluids | SUS 316L, PEEK, FKM |
Steam sterilization | 140°C / 0.6 MPa, within 60 minutes |
Connection bore diameter | IDF / ISO 1.5 S ferrule |
Weight | Approx. 0.3 kg |