Conductivity meter (For high concentrations) CEH-200

Overview

With the purpose of attaining highly accurate measurement at the worksite, this design fulfills basic specifications while raising the bar in terms of ease of operation. Platinum temperature measuring resistor is utilized in temperature compensation element thereby allowing realization of high accuracy temperature compensation. The measurement of salinity of salt water is also possible.

Features

  • Utilizes 5 alternating current electrode sensors
  • High accuracy temperature compensation
  • Automatic determination of sensor quality
  • Continuous measurement of saltwater salinity and temperature is possible
  • Utilizes large customized LCD
  • Measured values and parameter settings displayed simultaneously

Manufactured by HORIBA Advanced Techno

Specifications

Format

CEH-200

Measurement range

Electrode format (4-electrode system)

Temperature

1.00 / 2.00 / 5.00 / 10.0 / 20.0 mS / cm (Conductivity mS / cm display)
1000 / 2000 / 5000 µS / cm (Conductivity mS / cm display)
0 - 4.0 % (Salinity display)
0 - 100 °C (Temperature display Measurement only)

Reproducibility

± 0.05 % / FS

Transmission output

DC 4 - 20 mA Input / output isolated, Maximum load resistance 500 Ω Linearity 0.5 % FS

Transmission output range

1.00 / 2.00 / 5.00 / 10.0 / 20.0 mS / cm
1000 / 2000 / 5000 µS / cm 0.0 4.0 %
(Changeable with internal DIG switch)

Adjustment function

1c contact output 2CH independent capacity: AC 250 V, 3 A max (Resistance load)

Control operation

ON / OFF Control: Upper limit operation (H) and lower limit operation (L) (Changeable with internal DIP switch)

Upper / lower limit setting range

Within each measurement range

Control width

Each range full scale 0 - 20 % (0 - ± 10 %)
Settings can be specified as absolute values

Calibration function

Conductivity: Based on cell constant setting
Temperature: 1-point or 2-point calibration for known temperature

Additional function

Each error message type
1. Sensor diagnostics error (Temperature sensor short, Temperature sensor tripped)
2. Measurement out-of-range error (Full scale over / under)
3. System error (A/D converter defect, Memory anomaly) Transmission output during calibration Continuous, Hold (Previous value hold)
Temperature calibration function 1-point or 2-point calibration is possible

Temperature conversion coefficient arbitrary setting

0 - ± 5 % / °C

Standard temperature

25 °C (Arbitrary setting possible: 5 - 95 °C)

Temperature compensation coefficient

NaCl or 0 - ± 5 % / °C (Arbitrary setting possible)

Temperature compensation range

0 - 100 °C

Temperature compensation accuracy

± 1 % / FS

Temperature compensation element

Pt 1000 Ω (0 °C)

Ambient temperature & humidity

-5 - 50 °C, Less than 85 % RH

Power

AC 100 - 115 V 50 / 60 Hz 7 VA
or AC 200 - 240 V 50 / 60 Hz 7 VA
(Depends on setting)

Construction

Outside equipment: JIS C0920
Protection class: 3 (Rain-proof type)
Installation procedure: 50 A pole or wall surface installation
Case: Aluminum alloy casting (Paint color: DIC G-262 or equivalent)
Full cover: Polycarbonate (Resin color: DIC G-262 or equivalent)
Hood: SUS304 (Paint color: DIC G-262 or equivalent, or DIC PART2 2600 or equivalent)
Mount fittings: SUS304

External dimensions

222 (W) x 180 (H) x 155 (D) (Mount fittings and protrusion not included)

Weight

Approx. 4 kg

Related Products

Conductivity meter CE-200C

High accuracy temperature compensation CPU, simultaneous display of conductivity and temperature,...