K-8 series on-line pH/ORP meter
Overview
The HORIBA K-8 process pH (ORP) analysis system features a systematic combination of functional component units which offers the versatility needed to meet a wide range of applications.
Features
- A Sensor for All Samples Versatile HORIBA three-in-one sensors
- User Friendly, Trouble Free Submersible & flow-type sensor holders
- Trust your Transmitter
2-sire & AC-operated versions with fully isolated output Digital or analog display - The Power Behind the System
HORIBA power units offer a perfect match but any 24~30V DC power supply acceptable - A Quartet of Cleaner Systems
Four series to maintain high sensor accuracy
Manufactured by HORIBA
Specifications
Range:
[pH] 0 -14, 0 -10, 2 -12,
4 -14, 0-8, 3-11,
6-14 or 4-10pH
[ORP] 0-±700 or 0-1000 mV
Accuracy:
[pH] ±0.1 pH, analog
±0.05 pH, digital
[ORP] ±10 mV, analog
±5 mV, digital
Data Display:
[Analog] 200 mm scale, wide angle meter graduated in 0.2 pH or 20 mV
[Digital] 3 1/2 dig LCD character height; 17.5 mm
Output:
4-20 mA DC Isolated from electrode circuit
Total load resistance; 600 Ω at max.
Cable Extension:
Output signal; approx. 200 m at max.
Sensor cable; 100 m at max.*
* Use cable junction box JB-30 and HORIBA W-5B cable.
Sample Conditions:
Temperature; - 10 – 110℃**
** Select the sensor and the sensor holder.
Pressure; 0 - 0.3 kg/cm2G
Flow rate/velocity;
[Flow-type] 1 - 15R/min.
[Submersible] Less than 2 m/sec.
Ambient Temperature:
- 10 – 50 ℃
Sample Contacting Material:
[Submersible holder]
Polypropylene or stainless steel 316
[Flow-type holder]
PVC, polypropylene and/or stainless steel 316
Cable Connection:
For sensor cable; PF 1/2 or JIS A11 equivalent
For output signal and power supply; PF 1/2 or JIS A13 equivalent
Temperature Compensation:
Automatic compensation by integral posister in the sensor
Transmitter:
a) 2-wire transmitter, or b) AC-operated version
Weather-proof enclosure of polypropylene oxide (PPO),mountable on 2B pipe
Complete with an analog meter or digital readout
Power;
[2-wire transmitter]
24 V DC, approx. 0.5 VA obtained by HORIBA power unit described below, or commercially available power supply of 24 - 30 V DC.
[AC-operated version]
100 (85 - 115), 115 (95 - 135) or 220 (190 - 260) V AC as specified, 50/60 Hz
Power Unit:
Available in rack type, wall-mount or weather-proof unit
Output; 24 V DC, 4 - 20 or 0 - 16 mA DC
Power; 100(85 - 115), 115(95 - 135)
or 220 (190 - 260) V AC as specified, 50/60 Hz