- Download the LSH Illuminator Specification Sheet
Optical Specifications | |||||
---|---|---|---|---|---|
LSH-D | LSH-100 | LSH-T250 | LSH-GB | LSH-GC | |
Lamp type | 100 W Deuterium | 100 W Tungesten Halogen | 250 W Tungesten Halogen | 140 W Glow Bar | 22 W Ceramic |
Reflector | f/4 | f/4 | f/4 | f/4 | f/4 |
Broadband optical power at focal point | 0.001 W | 1 W | 2 W | 0.001 W | 0.005 W |
Nominal image size | 22 | 9.5 mm H x 6.2 mm W (flattened helix) | 11.7 mm H x 5.5 mm W (cylindrical helix) | 17.5 mm H x 6.4 mm Ø (cylinder) | 10 mm H x 3 mm Ø (cylinder) |
Color temp. at rated power | N/A | 3300 K | 3400 K | 1000 -1050 K | 16000 - 2000 K |
Focal distance | 3.62 inches, 92 mm, from front of housing | ||||
Housing dimensions | 8.44 x 6 x 5.19 inches, 214 x 152 x 132 mm (L x W x H) | ||||
Weight | 12 pounds, 5.45 kg | ||||
LPS-QTH Power Supply Specifications (for TH, GB & GC sources) | |||||
Constant voltage operation | Regulation | Line regulation ≤ 5 mV, Load regulation ≤ 5 mV | |||
Ripple & noise | ≤ 5 mV rms, 100 mV p-p 20Hz ~ 20 MHz | ||||
Recovery time | ≤ 500 µs (50% load change, minimum load 0.5 A) | ||||
Temp. coefficient | ≤ 100 ppm/°C | ||||
Output range | 0 to rating voltage continuously adjustable | ||||
Constant current operation | Regulation | Line regulation ≤ 3 mA, Load regulation ≤ 3 mA | |||
Ripple current | ≤ 10 mArms | ||||
Output range | 0 to rating current continuously adjustable (Hi/Lo range switchable) | ||||
Meter | Type: 3 1/2 Digit 0.39" LED Display, Accuracy: ±( 0.5% of rdg + 2 digits ) | ||||
Insulation | Chasis & terminal: 20 MΩ or above (DC 500V), Chasis & AC cord: 30 MΩ or above (DC 500V) | ||||
1684P Deuterium Power Supply Specifications (for Deuterium sources) | |||||
Input | Voltage | AC 100/240 V, 50/60 Hz | |||
Current (Max) | 0.9 A | ||||
Output | Voltage (DC) | 80 V (Typ.) with Load, 200 V (Min.) without Load | |||
Current (DC) | 300 +/- 30 mA | ||||
Current fluctuation (p-p) | 0.005% (Typ.) | ||||
Current drift at +25°C | +/- 0.02%/h (Typ.) | ||||
Warm-up time | 20 seconds (Approx.) | ||||
Trigger voltage | 600 V peak (Approx.) | ||||
Oper. ambient temp. | 0 to +40°C | ||||
Storage temperature | -10 to +60°C | ||||
Operating & storage humidity | Below 80% (no condensation) | ||||
Weight | 1.8 kg |