EMIA-V2 series Carbon/Sulfur Combustion Analyzer

Overview

The EMIA-V2 series measures carbon and sulfur extracted during combustion in a programmable High Frequency furnace without conversion. Combustion control is achieved through the use of the high-frequency induction furnace method. Appropriate heat-control in accordance with the analysis objective enables high-precision analysis for all types of samples. A variety of models allow you to select the furnace system that is best suited to your target samples and application.

  • EMIA-920V2 C/S analyzer: High performance analyzers with outstanding accuracy of less than 0.3ppm
  • EMIA-320V C/S analyzer: Standard performance analyzers for wide range of applications.

Features

  • "Auto cleaner" mechanism reduced maintenance works
    (Standard for 920V, Option for 320V2)

  • The Dust filter heating mechanism provides stable analysis with extremely low gas absorption
    (920V2 only)

  • Real-time simultaneous detection of CO/CO2/SO2 without an oxidizing (COCO2) or SO3 trapping reagents.

  • Wide variety of self diagnostics

  • Programmed temperature curve for optimizing combustion.

  • Choose from standard or wide range models to suit your application requirements.


Manufactured by HORIBA

Specifications

EMIA-920V2 series
EMIA-920V2:Simultaneous carbon/sulfur analysis
EMIA-921V2:
Carbon analysis
EMIA-922V2: Sulfur analysis

Principle:
NDIR/ Non Dispersive Infrared Gas Anaryzer
High-frequency furnace method in oxygen flow

Range: Carbon: 0-6,0%, Sulfur: 0-1,0%
Up to 100% wt is possible by decreasing the sample weight.

Accuracy
Carbon:
Concentration less than 20 ppm >>> σn-1≦0.3 ppm
Concentration higher than 20 ppm >>> σn-1≦1 ppm or RSD ≦0.5%

Sulfur:
Concentration less than 20 ppm >>> σn-1≦0.3 ppm
Concentration higher than 20 ppm >>> σn-1≦1 ppm or RSD ≦0.75%

Sample weight: 1,0g standard

Sensitivit: 0,01ppm

Analysis time: Normally 30 to 60 seconds(after combustion),varies according to the characteristics of the samples

Calibration
One-point or multi-point calibration using standard sample

Combustion furnace method/furnace structure
High-frequency induction furnace method with combustion control function (patent pending)
Anode output: 2300W
Frequency: 20MHz
Plate current: Settable up to 500mA 
Automatic control possible (10 slopes maximum)
Constant dust filter heating function

Operation gas
Oxygen purity: More than 99.5%
Dry air or nitrogen: Supply pressure 0.35MPa

Functions
Interrupt analysis
Self diagnosis
Data storage(Measurement results)
Traceability of calibration curve and measurement conditions
Statistic calculation
Maintenance navigator

Automatic cleaning mechanism: Built-in as standard

Oxygen gas purifying unit: Built-in as standard

Power
Main unit: AC200/220/240V  5kVA  50/60Hz
PC: AC100V, 120/200/220/240V  0.5kVA  50/60Hz

Dimensions: 560(W)×927(D)×950(H)mm Approx.170 kg
--------------------------------------------------------------------------------------------------------

EMIA-320V2 series
EMIA-320V2:Simultaneous carbon/sulfur analysis
EMIA-321V2:Carbon analysis
EMIA-322V2:Sulfur analysis

Principle:
NDIR/ Non Dispersive Infrared Gas Anaryzer
High-frequency furnace method in oxygen flow

Range:Carbon: 0-6.0%, Sulfur: 0-1.0%
Up to 100% wt is possible by decreasing the sample weight.

Accuracy
Carbon: σn-1≦2 ppm or RSD≦1.0%
Sulfur: σn-1≦2 ppm or RSD≦1.5%

Sample weight: 1,0g standard

Sensitivity: 0,1ppm

Analysis time
Normally 30 to 60 seconds(after combustion),varies according to the characteristics of the samples

Calibration: One-point or multi-point calibration using standard sample

Combustion furnace method/furnace structure
High-frequency induction furnace method with combustion control function (patent pending)
Anode output: 2300W
Frequency: 20MHz
Plate current: Settable up to 500mA 
Automatic control possible (10 steps maximum)

Operation gas
Oxygen purity: More than 99.5%
Dry air or nitrogen: Supply pressure 0.35MPa

Functions
Interrupt analysis
Self diagnosis
Data storage(Measurement results)
Memory of calibration curve and measurement conditions
Statistic calculation

Automatic cleaning mechanism: Option

Oxygen gas purifying unit: Option

Power
Main unit: AC200/220/240V  5kVA  50/60Hz
PC: AC100V, 120/200/220/240V  0,5kVA  50/60Hz

Dimensions: 560(W)×885(D)×820(H)mm Approx.170 kg

Schematics

EMIA-Vseries Dimensional outline drawing (mm)

Brochures

EMIA-V2

Related Products

EMIA-8100 series

EMIA-8100 series equipped with electrical resistance furnace measures carbon and sulfur contained...

EMGA-920

This is an oxygen and nitrogen elemental analyzer with high accuracy and repeatability suiting to...

EMGA-600W

The HORIBA EMGA-600 is a combined model between ON and H analyzers and determines the oxygen,...

SLFA-2100/2800

The SLFA-2100/2800 are designed specifically to meet the recent demanding needs of measuring the...

XGT-5200WR

High sensitivity XRF qualitative and quantitative analysis with an X-ray imaging...

XGT-7200

Single point analysis and automated hyperspectral imaging.

Dual vacuum modes.

Spot sizes from 1.2...

XGT-5200

Single point analysis and automated imaging.

Atmospheric pressure analysis.

Spot sizes from 1.2...

GD-Profiler

The GD-Profiler 2™ provides fast, simultaneous analysis of all elements of interest including the...

GD-Profiler HR

The HORIBA Scientific GD-Profiler HR™ gives the optimum in terms of resolution and number of...

Assistance tools with ICP-OES

ACTIVA-M, a bronze award winner for best new product at Pittcon 2007 as Master in multi-line...