Sygnature™ Linear CCD Detection System

Overview

Sygnature™ is a compact linear detection system offering outstanding sensitivity and acquisition speed. It is ideally suited for UV-VIS-NIR absorption, colorimetric, emission, and many fluorescence measurements. The Sygnature™ detectors come in two varieties, the 3648 pixel CCD and the 1024 pixel PDA. The 3648 pixel CCD has 8 µm x 200 µm pixels which provide uncompromised resolution over its 29 mm focal plane and highest sensitivity for lower signal levels. The 1024 pixel PDA is best for medium to high light level applications that can take advantage of its greater dynamic range.
The Sygnature™ detectors are designed to be interfaced to any HORIBA Jobin Yvon spectrograph and can provide an easy upgrade to existing SPEX, HORIBA Jobin Yvon, and ISA monochromators* and spectrographs.

A single USB 2.0 interface provides both high speed transfer and power. The basic software package, Lynear™, offers full exposure control and data transfer as well as control of I/O trigger functions. SygnatureT also runs under HORIBA Jobin Yvon's powerful SynerJY® software. When Sygnature™ is coupled with a MicroHR™ spectrograph, the combination provides better optical performance than self-contained linear CCD spectrometers, flexibility in spectral coverage, and unmatched versatility as a result of the wide choice of interchangeable gratings available for the MicroHR™.

* Additional adapters may be required for single port monochromator systems.

Features

  • Low Cost
  • Broad spectral coverage (between 200 nm and 1100nm)
  • Fast
  • Computer controlled Triggers
  • Truly plug and play
  • Multiple software options
  • Compact and portable
  • Easily mounts to most HORIBA Jobin Yvon spectrometers
Benefits
  • Less expensive than a PMT or silicon photodiode
  • Ideal for most spectroscopic applications
  • Acquires up to 160 spectra per second
  • Well suited to many time dependent experiments and kinetic studies
  • High speed USB 2.0 interface
  • Lynear™ basic acquisition software included, SynerJY®, single application packages, and LabVIEW® VIs also available
  • Flexible Spectrometer without taking up lab space
  • Gives new life to that old system sitting on the shelf

Manufactured by HORIBA Scientific

Specifications

Model Number

Sygnature-CCD-3648-VIS

Sygnature-PDA-1024-VIS

Sensor

Toshiba TCD 1304AP Linear CCD

Hamamatsu S3903-1024Q PDA

Pixels

3648 8 µm x 200 µm

1024 25 µm x 500 µm

Spectral Range

200 nm to 1100 nm

200 nm to 1100 nm

Dynamic Range

1250 : 1 (10+ bits)

6000 : 1 (12+ bits)

Spectral Acquisition Time

20 ms per spectrum

6 ms per spectrum

Integration Time

10 ms to 65 s

3 ms to 65 s

Input Trigger

TTL High Edge Trigger

Output Trigger/Shutter Control

TTL programmed to exposure time or dark subtraction

Computer Interface

USB 2.0 (USB 1.1 Compatible)

Included Software

Lynear™ basic data acquisition package
Optional: SynerJY® Spectroscopic Software

Operating System

Windows 2000/XP

Size

4.0 in f by 2.06 in (102 mm f by 52.3 mm)

Weight

0.8 lb (0.36 kg)

Specifications subject to change without notice.

Sygnature + MicroHR
  • Low Cost
  • High Performance
  • Extremely Versatile
  • Field Portable

Spectral Performance of a Sygnature™-3648 & MicroHR™ Package*

Grating
(Grooves/mm)

Coverage
(mm)

Typical Working Ranges

Resolution: 10µm slit
(nm FWHM)

Resolution: 100 µm fiber
(nm FWHM)

1200

153

High resolution

0.06

0.5

600

300

400 nm - 700 nm

0.12

1.0

300

613

200 nm - 800 nm

0.24

2.0

150

1200

200 nm - 1100 nm**

0.48

4.0

 * All measurements taken at 633 nm

** Actual array coverage is 200 nm to 1400 nm, however the detector is only sensitive to 1100 nm

Related Products