H20-UVL

Overview

VUV Monochromator

The H20-UVL is a monochromator especially designed for analyzing 100-400 nm (3 to 12.4 eV) Vacuum Ultraviolet (VUV) rangewhen using under vacuum, or 190-400 nm at atmospheric pressure. Its micrometric slits and its worm drive make its scansprecise and fast. This short focal length vacuum monochromator is ideal for lighting sample if couple with a VUV light source or forVUV low resolution analysis when equipped with single channel detection.

Based on HORIBA Jobin Yvon’s patented technology, the H20-UVL series is built around a single spherical holographic grating aberration corrected. Its 64° deviation angle makes this monochromator perfectly optimized in VUV range.

This simple optical design dramatically reduces astigmatism and results in excellent luminosity and spectral purity, even below 140 nm, where other instruments based on Czerny Turner design loose their efficiency because of the number of internal reflections.

Features

  • Single Grating design
  • Type IV Grating
  • MgF2 coating UV optimized
  • Low Astigmatism level
  • High Vacuum compatible
  • Automate drive
  • Build-in USB2 interfaces
  • HJY Slit Attachment

Benefits

  • Optimized for throughput
  • Minimized aberrations
  • Better efficiency in VUV range
  • High S/N ratio measurements
  • 10-5 mbar or optional 10-9 mbar
  • Fast and Easy to operate
  • No additional controller. Easily programmable with SDK.
  • Compatible with all HJY accessories

More Information


Manufactured by HORIBA Scientific

Specifications

Optical design

Spherical Type IV (single optic)

Focal length

200 mm

Aperture

f/4.2

Grating density

1200 gr/mm

Optic coating

MgF2 optimized at 121 nm

Deviation angle

64°

Dispersion

3.6 nm/mm at 100 nm

Drive

Fast worm drive

Minimum step

0.06 nm

Speed

400 nm/s

Accuracy

+/- 0.2 nm or +/-0.06 nm (*)

Repeatability

+/-0.06 nm

Resolution

Better than 0.5 nm (**)

High Vacuum

10-5 mbar (***)

Pumping flange

DN40 KF

Entrance/exit port

Micrometric slits (10 um to 2 mm)

Entrance/exit flange

DN25 KF

PC interface

USB2

Schematics

Applications

  • Transmission-Reflection measurements
  • UV tunable filter/Light source
  • Laser Harmonic Filter
  • Quality control in VUV

Accessories

Options

Ultra High vacuum

10-9 mbar

Slit height reducers

2, 4, 6 mm diameter

 

 


Related Pages