Dissolved Oxygen Monitor SD-300
Overview
Enables High-precision Measurement and Monitoring of DO in Ultra-pure Water in the Semiconductor Industry
Ever-increasing semiconductor circuit densities require higher-quality ultra-pure water for cleaning wafers.
Minute DO (dissolved oxygen) in ultra-pure water oxygenates the wafer surfaces. To ensure high-quality ultra-pure water, DO levels must be strictly controlled. Horiba's SD-300 DO Monitor is the solution.
Created utilizing Horiba's semiconductor technologies, it provides effective high-quality, pure-water management. The SD-300 DO Monitor is lightweight and portable. It can be carried anywhere, enabling high-precision measurement of DO levels in ultra-pure water in the semiconductor industry.
Features
- Portable Design for Easy Operation Even in Narrow Spaces
Compact, portable, and easy-to-use. - Achieve a Wide-range of Highly Sensitive Measurements
Select the measurement range: 0 to 20 µg / L, 0 to 200 µg / L, 0 to 2 mg / L, or 0 to 20 mg / L. Switch between automatic and manual operation. Achieve highly sensitive measurements in the optimum range. - High-speed Response
- Flat Panel Design for Use in Clean Rooms
- Durable Internal Liquid Replacement Sensor
- Standard-feature Analog / Serial Output
Manufactured by HORIBA Advanced Techno
Specifications
Measuring object | Dissolved oxygen | |
|---|---|---|
Principle | Membrane Polarography | |
Measurement range | 0 to 20 µg / L, 0 to 200 µg / L, 0 to 2 mg / L, 0 to 20 mg / L | |
Response time | 30 s / 90 % (at 20 °C or more) | |
Repeatability | ± 1 % of full scale | |
Temperature compensation | 5 to 50 °C | |
Sample conditions | Temperature | 2 to 50 °C |
Pressure | 0.02 to 0.2 Mpa | |
Flow rate | 0.2 to 0.4 L / min. | |
Output | Analog output | 4 to 20 mA DC or 0 to 16 mA DC, isolated output load: 500 Ω max. |
Contact output | Contact A output, contact capacity: 24 V DC, 0.1 A | |
Serial output | RS-232C output only | |
Calibration | Zero | Electrical calibration or calibration with zero solution |
Span | Air (by placing a sensor in the air) | |
Ambient condition | Temperature : 0 to 45 °C, Humidity : 85 % max. | |
Power | 100 to 240 V AC (± 10 %), 50 / 60 Hz | |
Power consumption | Approx. 50 VA (at 100 VAC) | |
Dimension | 296 (W) X 110 (H) X 258 (D) mm (operation unit only) | |
Mass | Approx. 3.4 kg (operation unit only) | |
Length of extension cable between | 5 m (standard), 10 m (option) | |