ABX Micros ES 60

Overview

  • Integrated barcode reader
  • USB connection
  • Unique pack system
  • Integrated ticket printer

Features

ABX Micros 60

  • RBC/PLT Detection Principles: Method Impedance, Ruby diameter 50 μm, Depression of count 200 mb, Duration of count 2 (or 3) x 6 seconds, Dilution ratio 1/15 000
  • HGB Measurement: Method Photometry, Wavelength 550 nm, Dilution ratio 1/250
  • HCT Measurement: Method Numeric integration
  • WBC Measurement: Method Impedance, Ruby diameter 80 μm, Depression of count 200 mb, Duration of count 2 (or 3) x 6 seconds, Dilution ratio 1/260
  • 3 Part Differentiation: Method Impedance + specific lyse action
  • MCV, MCH, MCHC, RDW, PCT*, PDW* Calculation
  • ABX Micro-sampling: Only 10μl of whole blood per analysis
  • ABX Liquid valves: Precise volume and reliability
  • ABX stepper motor vacuum pump: Noiseless running (no-compressor)

* RUO parameters (Research Use Only)


Manufactured by HORIBA ABX SAS

Specifications

Technical

Weight

16 kg - 30.9 lb

Power supply

100 V to 240 V (± 10%) 50 Hz to 60 Hz

Dimensions

43 x 36 x 36 cm (HxLxP) . 16.9 in 14.2 in 14.2

Throughput

OT: 60 samples/hour / CT: 50 samples/hour

 

Operating Temperature & Humidity

6 to 30°C (61 to 86°F) room temperature
Maximum relative humidity 85% for temperature up to 30°C (86°F)

Software

8.4” LCD screen / 256,000 colors, 640 x 480 pixel resolution
Touch screen technology / Inverter card for LCD screen
Secure digital memory card: 512 Mo / Linux Operating System

Certification

EN 61326: 2001
EN 61326-2-6: 2006
IEC 61010-1: 2001
IEC 61010-2-81: 2001 / A1: 2003
IEC 61010-2-101: 2002UL61010-1: 2004
C22.2 N°61010-1: 2004
In Vitro Diagnostics Directive: 98/79/EC
cTUVus mark

Connectivity

Hardware: 7 connections ports
ABX and ASTM protocols available
ML compatibility

Printer

58mm thermal printer with rapid paper loading

Samples

60 samples/hour (open tube)

Parameters (16+2)

CBC Mode (18 parameters): WBC, RBC, HGB, HCT, MCV,*RDW, MCH, MCHC, PLT, MPV, PCT, *PDW, % and # of LYM, MON and GRA

Graphics

LMG histogram, RBC, PLT histograms

Sample-Volume

CBC mode : 10 µL

Stability

24 hours (CBC)

Operating sample mode

Open tube mode with 4 position tube holder (QC, 125 or 500 µL, 3 or 5 mL)

Identification

Patient I.D., Sequence#

Flags

Pathological flags (WBC, RBC, and PLT) normal limits and counting failure flags

 * RUO (Research Use Only)

Results

Parameters

Presicion (cv %)

Linearity

WBC

< 2,5%

0-100 x 10³/mm3

RBC

< 2%

0-8 x 106/mm3

HGB

< 1.5%

0-26 g/dl

HCT

< 2%

0-80%

PLT (whole blood)

< 5%

0-2200 x 10³/mm3

PLT (PLT concentrate)

< 5%

0-4000 x 10³/mm3