ABX Pentra DF 120 SPS

Overview

  • Autoloader with continuous loading capability
  • Double DIFF Matrix
  • Automatic reflex testing
  • Validation station with cytology atlas
  • DHSS* technology (*HABX patent)

Features

ABX Pentra DF 120

  • Methods of measurement: Cytochemistry, impedance, absorbance and flow cytometry
  • Balance Concept: Automatic control of the leukocyte count based on 3 independent channels
  • DHSS: Focused flow cytometry and sequential measurement (impedance and absorbance)
  • Automatic re-run of samples: Selective and programmable (hematology parameters alarms and flags)
  • Rotative mixing of samples: Smooth and efficient
  • Liquid valves ABX (no pinch valves): Reliability and volume accuracy

SPS Evolution

  • Drop needle system: Auto-cleaning drop needle system. Calibration of blood drop.
  • Smearing system: No inter-sampling contamination, No carry-over, Easy to replace.
  • Stainer: Choise of Stain: MGG, Wright, Wright Giemsa. Programmable staining protocol.

Manufactured by HORIBA ABX SAS

Specifications

Technical

Weight

170 kg

Power supply

100/240 V - 50/60 Hz - 700 W
laser in stand-by 800 W, (2kW maxi).

Dimensions

75 x 167 x 55 (H x W x D) cm

Printer

Laser printer

Throughout

120 tests/hour

Data processing

OS9
Floppy disc, RS 232C, Mono and bi-directional connection

Compliance

ASTM protocol

Memory capacity

90 000 results + graphics

Samples

Parameters (32)

WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, PLT, MPV,PCT*, PDW*, NEU, LYM, MON, EOS, BAS, ALY, LIC, IMM, IMG, IML: (% and #)

Graphs

High-definition DIFF matrix, PLT, RBC, WBC/BAS curves

 

Volume

Manual mode : 130 µL
Automatic mode : 200 µL 

Slide

50 µL

Stability

Up to 48 hours for CBC/DIFF

Operating sample mode

Autoloader 15 racks (150 tubes) capacity
Sample mixing in end-over-end rotation mixing
Open tube in STAT mode

Identification

Alphanumeric, Bar-code reader (internal and external)
Automatic impression in indelible ink on itch slides (3 lines)

Results

Parameters

Presicion (cv %)

Linearity

WBC

< 2%

0-150 10³ / mm³

RBC

< 2%

0-8 106 / mm³

HGB

< 1%

0-24 g/dl

HCT

< 2%

0-67%

PLT (whole blood)

< 5%

0-1900 10³ / mm³

PLT (PLT concentrate)

< 5%

0-2800 10³ / mm³

LYM

< 5%

--

MON

< 15%

--

NEU

< 3%

--

EOS

< 25%

--

BAS

< 40%

--

Related Documents

Related Products

ABX Pentra DF 120

32 parameters

Expert validation system

ABX Pentra DX 120

49 parameters

Comprehensive validation system

ABX Pentra DX 120 SPS

49 parameters

Integrated slidemaker-stainer

Comprehensive validation system

ABX Pentra ML

Connect up to 3 instruments to the same validation station simultaneously

All-inclusive management...