ABX Pentra DX 120 SPS

Overview

  • Autoloader with continuous loading capability
  • Double DIFF Matrix
  • Automatic reflex testing
  • Automatic validation of results
  • DHSS technology

Features

ABX Pentra DX 120

  • Methods of measurement: Flow cytometry, Cytochemistry, impedance, absorbance, optical cytometry and fluoro-flowcytometry
  • QUADIMENSION Concept: Automatic correction of the leukocyte count based on 4 independent principles
  • Automatic reflex testing: Selective and programmable (hematology parameters alarms and flags)
  • Rack rotation mixing of samples: Smooth and efficient
  • Liquid valves (no pinch valves): Reliability and increased precision

SPS Evolution

  • Drop needle system: Auto-cleaning drop needle system. Calibration of blood drop.
  • Smearing system: No inter-sampling contamination, No carry-over, Easy to replace.
  • Programmable staining protocols: MGG, Wright, Wright Giemsa

Manufactured by HORIBA ABX SAS

Specifications

Technical

Weight

374 lb

Power supply

100/240 V - 50/60 Hz - 700W
laser in stand-by 800 W, (2kW maxi).

Dimensions

29.5 in x 65 in x 22.6 in (H x W x D)

Printer

Laser

Throughout

120 samples/hour

Data processing

  • Windows XP, Floppy disk,
  • CDROM (R/W), RS 232C, TCP/IP,
    2 x USB1 Mono and hi-directional connection

Compliance

ASTM

Memory capacity

80 GB unlimited patient database

Patient Management

Automatic validation (setting rules), Data Check, Patient history (matrix, curves, data), manual entry.

Samples

Parameters (49)

 WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, PLT, MPV, *PCT, *PDW %, NEU, LYM, MON, EOS, BAS, *ALY, *LIC, *IMM, *IML, *IMG : (% and #), RET (% and #), RETH %, RETM %, RETL %, *IMM %, CRC%, IRF %, MRV, MFI. NRBC % NRBC#, *CWBC, *WBC#, *RBC#, *PolyNuc#, *MonoNuc#
*RUO (Research Use Only)

Volume

Manual mode : 130 µL
Automatic mode : 200 µL 

Slide

50 uL automatic cycle

Stability

Up to 48 hours for CBC/DIFF

Operating sample mode

Autoloader 15 racks (150 tubes) capacity
Continuous feed for 8-racks, 160 slides capacity
Sample mixing with end-over-end rack rotation
Open tube in STAT mode

Identification

Alphanumeric, Bar-code reader (internal and external)
Automatic 3-line printing on the slide using indelible ink
Flags : 46 possible pathologic flags

Results

Parameters

Presicion (cv %)

Linearity

WBC

< 2%

0-150 x 103 / mm3

RBC

< 2%

0-8 x 106 / mm³

HGB

< 1%

0-24 g/dL

HCT

< 2%

0-67%

PLT (whole blood)

< 5%

0-1900 x 10³ / mm³

PLT (PLT concentrate)

< 5%

0-2800 x 10³ / mm³

LYM

< 5%

--

MON

< 15%

--

NEU

< 3%

--

EOS

< 25%

--

BAS

< 40%

--

Related Documents

Related Products

ABX Pentra DX 120

49 parameters

Comprehensive validation system

ABX Pentra DF 120

32 parameters

Expert validation system

ABX Pentra DF 120 SPS

32 parameters

integrated slide-marker