ABX Pentra XL 80

Overview

• CBC with 5-part Diff including flagging for immature cells
• Advanced technologies for mixing and sample analysis ensuring accurate results
• Processes up to 80 samples/hr and provides results in approximately 60 seconds
• Autoloader holds up to 100 samples with random continuous access
• STAT mode for critical samples

Features

• 20 parameters
• Whole blood micro-sampling (CBC: 35 µL, Plus Diff 53 µL)
• Cytochemistry, Impedance (real cell volume measurement) & Optical (analysis of the internal cellular structure by measuring light absorbance) - DHSS* Technology
• Perfect homogenization of blood samples with reagents - MDSS** Technology
• Automatic Sample Re-run
• Reagents: Only 4 reagents plus cleaner
• Perfect differentiation of the 5 WBC sub-populations with DHSS* Technology
• 3 histograms for RBC, BAS/WBC and PLT together with the 5 DIFF Matrix
• Basophils counted through specific channel
• High resolution matrix includes the determination of 2 additional subpopulations (% and #): Atypical Lymphocytes (ALY***) and Large Immature Cells (LIC***)
• Customized Dilution Ratio (CDR)
• Integrated Workstation and Validation Station 
* DHSS : Double Hydrodynamic Sequential System
** MDSS : Multi-Distribution Sampling System
*** RUO parameters (Research Use Only)


Manufactured by HORIBA ABX SAS

Specifications

Technical

Weight

122 lb / 55 kg

Power supply

100 V to 240 V (± 10%) - 50 Hz to 60 Hz

Dimensions

32.3 x 22.4 x 21.5 in / 82 cm x 57 cm x 54 cm.

Printer

Digital Laser

Throughput

Up to 80 samples/hour

Data Processing

Color LCD touch screen: 12 in

PC onboard: Celeron 566 MHz

Windows XP Prof.TM

RAM (256 MB), Hard disk (10 GB mini)

Floppy disk & CD ROM reader

RS 232C, 2 X USB1

User defined flagging limits

Transmit patient files & QC to LIS

Uni-directional & bi-directional connections

ASTM protocol inside


Memory capacity

10 000 results + Graphics

Quality Control Management

24 selectable QC files

XB: 60 operator selectable files with statistics (20 results per file)

Within run

Levey-Jennings graphs


Logs

Reagents, Quality Control, Calibration, Maintenance, Blank Cycle, Patients Data, Settings, Information, Technology Communication, Errors, by date.



Patient reports Management

Delta check

Archived results (Matrix, curves, data)

Manual input



 

Samples

80 samples/hour (open or closed tube)

Parameters (26)

WBC, NEU# & NEU%, LYM# & LYM%, MON# & MON%, EOS# & EOS%, BAS# & BAS%, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, PLT, MPV 

Graphs

High-definition LMNE matrix
PLT, RBC, WBC/BAS curves

Blood Volume

CBC mode: 35 µL
CBC + DIFF mode: 53 µL

Stability

48 hours post-draw stability

Operating sample mode

Autoloader 10 racks (100 tubes) capacity
Sample mixing with end-over-end tube rotation
Closed or open STAT tube

Identification

Alphanumeric. Bar-code reader (optional) and internal bar-code scanner

 

Results

Parameters

Precision (%CV)

Linearity Limits

Linearity Limits with CDR*

WBC

< 2.0

0-120 x 10³/µL

120-360 x 10³/µL

RBC

< 2.0

0-8 106/µL

0-8 X 106/µL

HGB

< 1.0

0-24 g/dl

0-24 g/dL

HCT

< 2.0

0-67%

0-67%

PLT (whole blood)

< 5.0

0-1900 x 10³/µL

1900-3800 x 10³/µL

PLT (PLT concentrate)

< 5.0

0-2800 x 10³/µL

2800-5600 x 10³/µL

*CDR: Customized Dilution Ratio