ABX Micros CRP 200

Overview

Simultaneous CBC & CRP

  • Single tube operation : Whole blood (EDTA)
  • Single microsample : CBC – 10uL; CBC + CRP - 18 µL
  • Fast and efficient process : No centrifugation
  • Inflammation testing : CBC and CRP
  • Compact analyzer
  • Integrated reagents
  • Reliable results
  • "Zero maintenance" compact data management impedance technology, photometry, immunoturbidimetry

Features

ABX Micros CRP 200

  • WBC, RBC and PLT : Impedance
    HGB : Spectrophotometry (Cyanmethemoglobin method)
    HCT : Numeric integration
    CRP : Latex immuno turbidity rate method
    MCV, MCH, MCHC, RDW, MPV, PCT* and PDW*.
    Calculations.

*: RUO parameters (Research Use Only)


Manufactured by HORIBA

Specifications

Technical

Power requirement

Power supply from 100 Vac to 240 Vac +/- 10%

  • 50 Hz to 60 Hz.
  • Power consumption Maximum 230 VA.
  • Operating Temperature and Humidity : 18°C - 30°C (64°F-86°F) room temperature.
  • Maximum relative humidity 80% for temperature up to 31°C (88°F) decreasing linearly to 50% relative humidity at 40°C (104°F).

Dimensions and Weight

Weight : 40 lbs.
Dimensions : W 11.8 x H 16.7 x D 15.7 in

Memory

QC (CBC) and QC (CRP) modes: 30 samples
CBC and CBC+CRP modes: 50 samples

Samples

Operational  mode

CBC mode 16 parameters with graphics for RBC, PLT and WBC populations.

CBC & CRP mode 16 parameters plus CRP

Sample-Volume

CBC mode : 10 µl
CBC & CRP mode : 18 µl

Parameters

WBC, RBC, HGB, HCT, PLT, CRP (C-reactive protein) MCV, MCH, MCHC, RDW, MPV, LYM%, LYM#, MON%, MON#, GRA%, GRA#, PCT*, PDW*.

Sample processing

On open tube with 4 position tube holder (QC, 500 µL, 3 mL, 5 mL)

Throughput analyses

CBC mode: Within 1 min, 15-sec (48 samples/hour)
CBC + CRP mode: Within 4min, 30-sec (14 samples/hour)

*: RUO parameters (Research Use Only)

Results

Linearity

Parameters

Linearity

WBC

        0-80 x 10³/mm3

RBC

        0-7.5 x 106/mm3

HGB

        0-23 g/dL

HCT

        0-62.4%

PLT

        0-900 x 10³/mm3

CRP (Whole Blood)

        2-200 mg/L

CRP (Serum)
(To measure up to 20 mg/L,
Hct value of sample must be > 23%.)

        2-150 mg/L

Precision

Parameters

Precision (CV %)

Range

WBC

< 2.5%

4.75 - 12.3 x 103/mm3

RBC

< 2%

4.0 - 5.0 x 106/mm3

HGB

< 1.7%

12.0 - 16.6 g/dL

HCT

< 2%

37 - 50%

PLT

< 5%

150 - 355 x 103/mm3

CRP

< 10%

0-10 mg/L

< 5%

> 10 mg/L

Related Documents

Related Products

ABX Micros 60

8 or 18 fully automated parameters

CBC / 3 Part DIFF

Open or Closed Tube