INM-700 In-Situ NOx/O2 Monitor
Product Overview
HORIBA's INM-700 in-situ gas analyzer is designed to meet the requirements of coal-fired power generation plants for cost-effective, low maintenance NOX monitoring. The analyzer uses a ceramic diffusion filter and Zirconium Oxide sensor directly in the gas stream, which eliminates the need for an extraction process for sample collection, and results in lower maintenance. The INM-700 is ideal for inlet SCR control, NOX measurements, and burner balancing/combustion tuning in boiler plant operations.
The INM-700 has the advantage of lower installation, maintenance, and capital costs, and a faster response time. In addition, for compliance strategies, the ceramic filter allows measurements to be taken near the combustion point of the furnace. This technique utilizes combustion modification techniques such as:
- Low NOX burners (LNB)
- Over-fire air injection (OFA)
- Gas re-burn systems
- Boiler tuning
- Combustion optimization programs
Applications
- SCR control
- Inlet/outlet NOX measurement
- Replacement of FGAS shelters
- Burner balancing/combustion tuning
- Improved burner efficiencies and reduction of NOX
- Identifying problem areas in the furnace due to gas stratification linked to specific burner columns.
Product Features
- A NOX sensor using a new technology for coal fired applications which can improve control of the ammonia injection rate in SCR systems.
- A practical NOX sensor with a fast response time that can be directly installed in the combustion gas stream.
- Extremely robust sensors that are long lived and capable of withstanding elevated temperatures.
Manufactured by HORIBA
Product Specifications
Model | INM-700 |
Principle | Zirconia type |
Measurement range | Nox: 0 to 100 ppm(minimum), |
Repeatability | Nox: ±1.0% of F.S. |
Drift | ±2.0% of F.S./week |
Response time(T90) | 10 sec. or less |
Display | LED digital display (effective 4 digits) for measuring value |
Output signal | Nox: DC 4 to 20 mA(maximum load 600 ) |
Calibration | Probe calibration (Automatic or manual) |
Warm up time | approx. 5 min |
Sensor cable | Max. 50m (164ft.)*1 |
Location | Indoor or outdoor |
Operational temperature | - 10 to 50°C(23 to 122°F) |
Humidity | 90% RH(no moisture condensation) |
Power | AC100V to 240V (50/60Hz) |
Power consumption | approx. 60VA |
Mass | Probe unit: approx. 10 kg (approx. 22lbs) |
Cooling air consumption: | 400°C (752°F): approx. 70 L/min.(2.47 cfm) |
Purge air | Pressure air: about 300kPa |