XE Analyzer - Multi-Gas CEMS

Overview

The HORIBA XE analyzer is designed for tomorrow's CEMS applications. The XE combines gas analysis, I/O for system controls, alarm functions, data storage, diagnostics, and a menu-driven user interface within a single compact case.

The front panel multi-line LCD display is menu driven, allowing the user to easily view a series of screens as well as perform extensive setup and configuration.

The instrument also incorporates a microprocessor and a flash card memory that allows for local data storage and facilitates software configuration updates.

Data communication and storage capabilities eliminate the need for programmable logic controllers - even for the most extensive data gathering requirement, such as 40 CFR 75 applications.

XE multi-gas CEMS literature

Features

  • Measurement of up to five gases.
  • Highly reliable with technology proven in thousands of installations.
  • No optical alignment required.
  • Field-proven technologies.
  • No routine maintenance required.

Manufactured by HORIBA

Specifications

Specification

Specification value

Measured components

One to five of the following gases: NOX, SO2, CO, CO2, and O2

Ranges Available

NOX

0-100/5000 ppm

CO

0-50/5000 ppm, standard or 0-10/5000 ppm

SO2

0-50/5000 ppm

NOX (CLD option)

0-10/5000 ppm

SO2 (UV OPTION)

0-10/5000 ppm

CO2

0-5/50 %

O2

0-10/25 %

Repeatability

Ranges >/= 200 ppm

± 0.5% of full scale

Ranges </= 200 ppm and O2 analyzer

± 1.0% of full scale

Response time

NOX, CO and CO2

15 seconds for 90% response

SO2 and O2

30 seconds for 90% response

Drift

Zero drift:

Ranges > 200 ppm

± 1.0% of full scale / 7 days

Ranges < 200 ppm

± 2.0% of full scale / 7 days

Span drift

± 2.0% of full scale / 7 days

Linearity

± 1.0% of full scale

Interference

<± 2.0% of full scale for standard sample gas conditions.

Power requirements

115 V AC, 50 / 60 Hz

Power consumption

300 VA for base configuration

600 VA for configuration with UV SO2 or CLD NOX options

Sample gas flow rate

0.5 lpm

Sample gas pressure

4.9 kPa (500 mm Aq) ± 0.5%

Outputs

0-16 mA or 4-20 mA and 0-1 V DC isolated outputs for up to 16 channels and 64 discrete digital outputs available ( all user selectable from list of available outputs)

Inputs

13 channels of analog 4-20 mA or 0-1 V DC and 8 discrete digital inputs standard, expandable as required.

Materials exposed to sample gas

304 stainless steel, TeflonTM, polupropylene, fluororubber, PVC, and glass

Display

320 X 240 dot matrix LCD

Dimensions;

Height:

5 ¼ in (134 mm)

Width:

19 in (483 mm)

Length:

22 in (559 mm)

Weight

Approx. 36 lb (16.3 kg)