XS Analyzer - Multi-Gas CEMS

XS Analyzer - Multi-Gas CEMS

Product Overview

The HORIBA XS analyzer utilizes chemiluminescence and magnetopneumatic as well as cross-flow modulated no-dispersive infrared (NDIR) technologies, providing highly accurate measurement for up to five gasses plus system control, yet requiring no routine maintenance nor optical alignment.

The XS analyzer features specialized designs such as the optimized path length and high sensitivity detector that give precision measurements with a sample flow rate of less than one liter per minute.

The XS analyzer automatically initiates system calibration at intervals predefined by the user. The easy-to-read display can show concentrations for any of up to five gasses, or display averages or diluent-corrected values.

A micro-processor provides true computerized control plus self-diagnostics that allow the user to easily determine the source of an alarm or caution signal displayed on the front panel.

Product Features

  • Measurement of up to five gases.
  • Highly accurate measurements.
  • No optical alignment required.
  • Field-proven technologies.
  • No routine maintenance required.

Manufactured by HORIBA

Product Specifications

SpecificationSpecification value

Measured Components

One to five of the following gases:

NOx, SO2, CO, CO2 and O2

Ranges available

NOX

CO

SO2

CO2

O2

 

0-100/5000 ppm

0-50/5000 ppm standard or 0-10/1000 ppm optional

0-50/5000 ppm standard

0-5/50 %

0-10/25 %

Repeatability

Ranges>/= 200 ppm

Ranges</= 200 ppm

 

± 0.5% of full scale

± 1.0% of full scale

Response time

NO, CO, and CO2

SO2 and O2

 

15 seconds for 90% response

30 seconds for 90% response

Drift

Zero drift

Ranges> 200 ppm

Ranges< 200 ppm

Span drift

For ambient temperature within ±10°F (5°C)

 

±1.0% of full scale / 7 days

±2.0% of full scale / 7 days

±2.0% of full scale / 7 days

Linearity

±1.0% of full scale

Interference

<±2.0% of full scale for standard sample gas composition

Power requirements

115 v ac, 50 / 60 Hz

Power consumption

300 VA

Sample gas flow rate

0.5 lpm

Sample gas pressure

4.9 kPa (500 mm Aq) ±0.5%

Outputs

0-16 mA or 4-20 mA and 0-1 V DCisolated output for up to 10 channels and 16 discrete outputs available rated 1A, 250V AC or 125 V DC resistive load.

Materials exposed to sample gas

304 stainless steel TeflonTM, polypropylene, fluororubber, PVC, and glass

Dimensions:

Height:

Width:

Length:

 

5 ¼ in (134 mm)

19 in (483 mm)

20 in (508 mm)

Weight

Approx. 34 lb (15.4 kg)