IT-540 Series Non-Contact Infrared Thermometers
Overview
The hand-held IT-540 Series of non-contact thermometers include six models. The various options offer versatility in measurement ranges, target areas and target distances as well as choices to include a target marker or without a target marker for all budgets. The non-contact thermometer allows for temperature measurement in dangerous applications such as monitoring abnormal heat generation in high-temperature voltage transformer equipment, power supply panels, quick and accurate hygienic monitoring of the storage temperature of food products. Measurements are instantaneous and accurate.
Infrared Thermometers product literature
IT-540 Series Instruction Manual
IT-540NH, N, S, E Instruction Manual
Features
Light, compact design with audible and visible alarm output. LED target marker shows temperature measurement point. Temperature Measurement range: -50 to 1000°C. Light weight and compact for portability. Small target size for accurate point measurements. Reasonable price. High accuracy.
- IT-540S
S-type(Spot focus model)
Uses an LED spot marker to allow quick measurement of electronic components with very small surface areas. - IT-540E
E-type(Standard model)
This model features high performance and an extensive range of functions at an extremely reasonable price. - IT-540N
N-type(narrow focus model)
Uses an LED twin beam marker with temperature alarms and a measurement hold function. - IT-540NH
NH-type(narrow focus model)
Uses an LED twin beam marker with temperature alarms and a measurement hold function. For temperatures to 1000°C - IT-540W
W-type(standard model)
Uses an LED spot marker to for 400mm distance measurements. - IT-540WE
WE-type(Standard model)
For 400 mm distances. Does not include a marker but it is the most economical non-contact thermometer.
Manufactured by HORIBA
Specifications
Measurement temp range: | Models IT-540N, S, E, W, WE: -50 to 500°C(-58 to 932°F) Model IT-540NH: -50 to 1000°C |
Target size: | Ø40/500mm(E, NH,N Type) ,Ø2.5/30mm(S Type),Ø100/400mm(W, WE Type) |
Detector/Optical lens: | Thermopile /Silicon |
Spectral response | Models NH, N, S, E: 8 to 16µm Models W, WE: 8 to 25µm |
Response time: | Less than 1.5sec (95% response) |
Accuracy: | 201 to 500°C ±1% of reading, 0 to 200°c ±2°C of reading, -50 to -1°C ±(-10% of reading +2)°C |
Repeatability: | ±1°C(-50 to -1°C: ±2°C) |
Sighting: | LED spot sight (S Type); 2-Beam laser sight (NH, N type); 1 beam laser sight (W type); None (E, WE type) |
Emissitivity setting: | 0.10 to 1.00 in 0.01 increments |
Alam setting: | Models N, S, E:-55 and 505°C(-67 to 941°F) Model NH: :-55 and 1010°C |
Automatic power OFF: | HOLD and SET mode: 15sec. |
|
|
Operating temp: | 0 to 50°C(E,S, WE Type); 0 to 40°C (NH, N, W Type) |
Operating humidity: | 35 to 85% RH, no condensation |
Storage temp: | 20 to 55°C no condensation |
Dimensions | 170(H) x 40(D) x 36(W)mm |
Weight: | (Batteries included) Approx.140g |