HO-200 (Four-Wire Analyzer)

Features

  • Redox potential measurement of various solutions
  • Suitable for aeration control of sewage treatment and neutralization of chemicals
  • Extensive interfaces (2-channel transmission output, RS-485 compatible, 4-relay output)

Manufactured by HORIBA Advanced Techno

Specifications

Measuring method

Metal electrode method

Measuring range

ORP: -2,000 to 2,000 mV   Resolution: 1 mV
Temperature: 0 to 100˚C   Resolution: 0.1˚C

Repeatability

ORP: ±5 mV or less   Temperature: ±0.3˚C (for equivalent input)

Linearity

ORP: ±5 mV or less   Temperature: ±0.3˚C (for equivalent input)

Transmission output

Two points 4 to 20 mA DC   Input/output insulated type   Maximum load resistance 900 Ω

Contact output

Three points Non-voltage contact output Relay contact, SPDT (1c)
 Contact function:
  R1, R2: Selectable from upper limit alarm, lower limit alarm, ON/OFF control, transmission output hold operation, cleaning output
  FAIL: Malfunction alarm
 Control operation: Control width: 2 to 400 mV (±1 to ±200 mV)

Contact input

One point   Contact format: Open collector non-voltage a contact
 Contact function: Cleaning operation external input

Communication function

RS-485   Two wire system, Input/output insulated type (not insulated from transmission output)

Ambient temperature

-20 to 55˚C

Temperature measurement element

Pt 1000 (0˚C)
Positive temperature sensitive resistor element: 500 Ω (25˚C), 6.8 kΩ (25˚C), 10 kΩ (25˚C)

Calibration method

Manual adjustment (offset) Correction (-200 to 200 mV)
Manual sensitivity correction (0.500 to 1.500)

Self-diagnosis function

Electrode diagnosis error, Analyzer malfunction

Power supply

90 to 264 V AC 50/60 Hz
Power consumption 15 W or less

Construction

Outdoor installation type: IP65 protection level
Mounting method: 50A pole or wall mounted
Case: Aluminum alloy
Mount fitting/hood: SUS304

Weight

Approx. 4.5 kg

Equipment protected

CE marking, FCC rules