Particle Characterization
SA-9600 Series Surface Area Analyzers
Overview
The SA-9600 series of surface area analyzers brings exceptional speed and convenience to BET surface area analysis. Now you can perform ultra-fast single-point surface area measurements or multi-point with push-button ease.
These tools use the flowing gas method to acquire gas adsorption and desorption curves. This information is then used to calculate total surface area according to the well known BET theory. The advantage of the flowing gas method is most evident in single-point mode as up to thirty sample analyses can be performed per hour.
The patented SA-9600 technology provides routine total surface area determination in as little as two to six minutes. The SA-9603 models feature three stations for three simultaneous surface area measurements.
Features
- Automatic Calibration
To ensure repeatable accuracy, the SA-9600 performs an automatic calibration before every measurement. And because it's fully automated, the SA-9600 technology eliminates variables otherwise introduced by operator involvement. - Lowest Cost of Ownership
The use of mass flow controllers in all versions of the SA-9600 series automatically create the necessary mixtures of nitrogen and helium for BET surface area analysis. This lowers the total cost of ownership by eliminating the need to purchase expensive gas mixtures. Additionally, service costs are minimal compared to static volumetric technology and the SA-9600 doesn't require difficult calibration. - Software Control
- The SA-9600 software was designed to be easy as 1,2,3
- Type sample name
- Choose measurement type
- Click "Start"
- Simply follow the three steps above and in a few minutes the full surface area report will be ready for the operator.
- View the measurement in real-time to see the auto-calibration, adsorption, and desorption.
- Flexibility
Functions allow optimization of system for different sample types. - Input exact concentration of mixed gas to choose the correct concentration of adsorbate to give results identical to the multipoint method.
- Ability to use many different adsorbates in addition to nitrogen, such as, argon, krypton, carbon monoxide, carbon dioxide and other non-corrosive gases.
- A wide variety of sample cells are available, allowing measurements to be made on diverse sample types.
- Use the feature-rich SA-9600 software to control the unit via USB communication, or use the built-in keypad and display.
- High Throughput
Every SA-9600 model includes multiple sample preparation stations (2 or 3 depending on model). For very high throughput environments the SA-9660 accessory provide three additional stations.
Manufactured by HORIBA
Specifications
- Surface Area Measurement Range
Total Surface Area: 0.1 to 50 m2
Specific Surface Area: Approximately 0.01 – 2,000 m2/g
Much more sensitive to low surface area samples than volumetric type analyzers, allowing the use of samples less than 1g and as low as 0.1 square meters in the sample cell. - Accuracy and Precision
Reproducibility is better than 0.5% COV.
Accuracy is better than 1% for most samples.
The SA-9600 Series offers a full line of high-quality, high-performance Surface Area Analyzers with four fully automatic analyzers to meet the needs of any research or quality assurance laboratory. The SA-9600 Series analyzers include:
- SA-9601: One sample analysis station, two sample preparation stations.
- SA-9601MP: One sample analysis station, two sample preparation stations, MFC gas mixture control and multi-point analysis.
- SA-9603: Three sample analysis stations, three sample preparation stations.
- SA-9603MP: Three sample analysis stations, three sample preparation stations, MFC gas mixture control and multi-point analysis.
- SA-9660: Three sample preparation stations.