Automated Blood Cell Counter with CRP measurement LC-178CRP

Overview

CBC CRP Simultaneous Measurement
PLUS
High-Specification Functionalities
Precision measurement of high-density test samples with CRP measurements (up to 20 mg/dL)

Features

  • Peripheral Printer included
  • ONE SYSTEM, DUAL PURPOSE
    Simultaneous measurement of CBC and CRP within the same sample.
  • COST REDUCTION
    Instrument price, test reagent costs, and maintenance expenses all reduced.
  • Complete measurement within 270 seconds
    HIGH-SPEED
    Elevated CRP results supports early detection of inflammation and infection.
  • COMPACT DESIGN
    5 different test reagents can be stored within the instrument.
  • INCREASED MEASURING RANGE
    CRP can be measured up to 20 mg/dL.
  • Electronic LOW-NOISE DESIGN
  • A maximum of 50 CBC + CRP results can be saved and printed from memory.
  • A maximum of 30 CBC + CRP (Quality Control) results can be saved and printed from memory.
  • SIMPLE MEASUREMENT WITH NO “PRE-PROCESSING” NEEDED
    Multi-tube sample holder Perfect for emergency testing of various size sample containers.
  • Safe for Micro sample containers
    WHOLE BLOOD, LOW VOLUME
    Easy measurement with small 18 μL samples

ONE SYSTEM, DUAL PURPOSE;

Two measurements are taken within the same sample.
Blood cell counts and CRP concentration measurements, which used to require separate test samples and methods, can now be carried out simultaneously using the LC-178CRP on just a single blood sample. In addition, saving time and reducing costs (reagent costs, maintenance expenses, etc.), makes it possible to minimize the use of disposable containers.

CBC (WBC 3-PART DIFFERENTIAL) AND CRP CONCENTRATION MEASUREMENTS

The LC-178CRP can simultaneously measure up to 19 different parameters for CBC and CRP concentration. The White blood cell count is a direct measurement, which also gives a three-part differential measurement of (lymphocytes, monocytes, and granulocytes) percentage and absolute values. With the ability of combining CRP measurement to a complete blood count, diagnosis and assessment of an illness will be more precise. The LC-178CRP is perfect for urgent testing and offers major benefits for conditions that require progressive observation, such as infection and inflammation.

WHOLE BLOOD, SMALL SAMPLE VOLUME

The LC-178CRP is capable of using a sample volume of 18μL of whole blood for CBC + CRP analysis and (10 μL of sample for CBC only). This small amount sample volume makes it possible perform a micro-sample extractions using micro-capillary and micro-container methods. This is an optimal solution for acquiring blood samples from children or from individuals who are difficult to extract normal venous blood from.
IMPORTANT: When using micro samples for analysis, collection techniques must be followed exactly as stated by the manufacturer. WBC three part differential and platelets may be inaccurately measured due to improper collection technique.

PREPROCESSING SAMPLES ARE NOT REQUIRED. OPTIMAL FOR EMERGENCIES

With the LC-178CRP, sample preparation, such as centrifugal separation and dilution before CRP measurement, as well as temperature adjustment, is not needed. Since measurement is possible immediately after collection of the blood sample, the LC-178CRP is perfect for emergency testing situations. Since the measurement operation is simple and easy, there is no need for a specialized staff for sample processing.

FAST and ACCURATE MEASUREMENTS [CBC+CRP mode: approximately 270 sec., CBC mode: approximately 75 sec.]

Since the LC-178CRP can complete its measurements within a relatively short time frame, accurate treatment can be performed on the spot.
Emergency response is greatest especially with children in whom conditions such as inflammation and infection can advance quickly and require urgent countermeasures.

INCREASED MEASUREMENT RANGE

Due to the development of new reagents, the LC-178CRP measurement range for CRP has been expanded. Measurement of high concentration samples of up to 20 mg/dL is now possible. Samples, which previously needed to be diluted, can now be measured accurately in a single analysis cycle, without dilution.
ImageImage


Manufactured by HORIBA

Specifications

Product name:

LC-178CRP Automated Blood Cell Counter with CRP measurement

Measurement modes:

CBC mode (Complete blood count parameters only), CBC+CRP mode
(complete blood count + CRP parameters), Quality Control (CBC) mode
(with statistical calculations) and Quality Control (CRP) mode (with statistical calculations)

Measurement parameters:

WBC, LYM%, LYM#, MON%, MON#, GRA%, GRA#, CRP, RBC, Hgb, Hct, MCV, MCH, MCHC, RDW, PLT, PDW, PCT and MPV
WBC 3-part differential distribution curve, RBC distribution curve and PLT distribution curve.

Measurement principles:

WBC, RBC, Hct and PLT: Electrical impedance method
Hgb: Cyanmethemoglobin method
CRP: Latex immunoturbidimetry (with Hct value compensation function)

Sample volume:

CBC mode and QC (CBC) mode: 10 L,
CBC+CRP mode and QC (CRP) mode: 18 L,

Measurement times:

CBC mode and QC(CBC) mode: approximately 75 sec.,
CBC+CRP mode and QC(CRP) mode : approximately 270 sec.

Sample tube holder:

For vacuum sample tubes, microsample tubes, control blood, and small sample tubes

Reagent storage:

Refrigerated to 15°C (CRP measurement reagents R2 and R3)

Display indicators:

Sample No. (an error code is displayed when an error is generated)Measured value display (WBC, RBC, Hgb, Hct, PLT, CRP)Status display (measuring, waiting for operator entry, processing, error)

Printout:

From peripheral printer, RS-232C connection
Measurement dates and times, sample No., error messages

Reprint:

Calibration factors, measurement results displayed on the test result display

Measurement conditions:

Ambient temperature: 18—30°C, relative humidity: 25—80% (without condensation)

Protection level:

Class 1 apparatus, with no attachments (JIS T 0601-1)

Power requirements:

100 V~240VAC, 50/60 Hz

Power consumption:

150 VA or less

Dimensions:

300 x 400 x 410 mm (W x D x H)

Weight:

Approximately 18 kg

Sample memory storage capacity:

CBC and CBC+CRP modes: 50 results maximum,
QC (CBC) or QC (CRP) mode: 30 results maximum