Chlorfenapyr

Overview

Chlorfenapyr is an insecticide and is used for multiple farm products because of its wide insecticidal activity range.  This kit allows an easy, rapid (within 2 hours), sensitive, quantitative, multi-sample analysis of chlorfenapyr that remains in farm products.  This kit is best applied for inspection at delivery time/prior to processing or primary screening in production sites.

 

 

 

Features

          Comparison with instrumental analysis

While instrumental analysis offers high accuracy, it requires complicated and time-consuming pretreatment techniques,and sophisticated instrument operation as well as  large quantities of hexane or diethyl ether as organic solvents.  The simpler pretreatment required for kit saves significant time.  Additionally, only a small quantity of methanol is required as an extraction solvent for analyzing chlorfenapyr.

          Application examples 

The figures show the comparison between standard curves with 10% methanol and reaction curves drawn with methanol extract of chlorofenapyr-added farm products (diluted to 10% methanol equivalent).  The overlapping curves; demonstrate the suitability of this kit for analyzing trace chlorofenapyr without any interference from the original produce. 


Manufactured by HORIBA

Specifications

The following data represent the basic kit performance. For more detailed specifications, please contact us. 

          Standard curve

          Cross-reactivity

Pesticides

Cross-reactivity (%)

Chlorfenapyr

100

Fludioxonil

< 0.1

Fluacrypyrim

< 0.1

Acetamiprid

< 0.1

Imidacloprid

< 0.1

Captan

< 0.1

Diazinon

< 0.1

Chlorpyrifos

< 0.1

Malathion

< 0.1

Etofenprox

< 0.1

Instrumental analysis (tomato)

Data courtesy of Dr. Koichi Shiraki, Gifu Prefectural Research Institute for Health and Environmental Sciences 

 

          Within-run reproducibility (tomato)

Samples

Average (ppb)

SD

CV(%)

A

4.3

0.2

3.6

B

8.6

0.5

6.3

n=8

         Between-analysts reproducibility (tomato)

Samples

Average (ppb)

SD

CV(%)

A

5.3

0.3

5.6

B

9.2

0.4

4.0

n=3

Daily reproducibility (tomato)

Samples

Average (ppb)

SD

CV(%)

A

4.3

0.0

0.7

B

8.8

0.8

8.6

n=3

         Recovery test (tomato)

Concentration of addition (ppm)

Recovery ratio (%)

4.0

126.5±8.4

1.0

118.5±5.8

n=3
Results of within-run reproducibility, daily reproducibility, between-analysis reproducibility and recovery test show that this test kit can quantitatively analyze chlorfenapyr in produce.

Main kit components / shelf-life

Contents (form)

Volume

Quantity

Antibody coated plate (dry form)

8 rows x 12 columns

1 piece

Standard reagent L (freeze-dried: 2ppb)

1mL

2 vials

Standard reagent H (freeze-dried: 10ppb)

1mL

2 vials

Enzyme-labeled Chlorfenapyr reagent (freeze-dried)

6mL (after dissolution)

2 vials

Washing reagent (solution: 10 times concentrated)

50mL

1 vial

Chromogenic reagent (solution)

13mL

1 vial

Stop reagent (solution)

13mL

1 vial

Shelf-life:

12 months after production

Storage at 2 to 8°C