H-Fenitrothion
Overview
Fenitrothion is an organic phosphorous insecticide. It is used for multiple farm products. This kit allows an easy, rapid (within 2 hours), sensitive, quantitative, multi-sample analysis of fenitrothion that remains in farm products. This kit is best applied for inspection af delivery time /prior to processing or primary screening in production sites.
Features
Comparison with instrumental analysis
While instrumental analysis offers high accuracy, it requires complicated and time-consuming pretreatment techniques, and sophisticated instrument operation as well as large quantities of hexane or acetone as organic solvents. The simpler pretreatment required for kit saves significant time. Additionally, only a small quantity of methanol is required as an extraction solvent for analyzing fenitrothion.
Application examples
The figures show the comparison between standard curves with 10% methanol and reaction curves drawn with methanol extract of fenitrothion-added farm products (diluted to 10% methanol equivalent). The overlapping curves; demonstrate the suitability of this kit for analyzing trace fenitrothion without any interference from the original produce.
Manufactured by HORIBA
Specifications
The following data represent the basic kit performance. For more detailed specifications, please contact us.
Standard curve
Cross-reactivity
Pesticides | Cross-reactivity (%) |
|---|---|
Fenitrothion | 100 |
EPN | 34 |
Parathion | 12 |
Parathion-methyl | 6.9 |
Fenitro oxon | 1.0 |
Cyanophos | 0.9 |
EPN Oxon | < 0.1 |
Fenthion | < 0.1 |
Diclofenthion | < 0.1 |
Diazinon | < 0.1 |
Isoxathion | < 0.1 |
Clorpirifos | < 0.1 |
Clorpirifos-methyl | < 0.1 |
Fenate | < 0.1 |
Chlorfenvinphos | < 0.1 |
Sulprofos | < 0.1 |
Protiophos | < 0.1 |
Propaphos | < 0.1 |
Profenophos | < 0.1 |
Within-run reproducibility (brown rice)
Samples | Average (ppb) | SD | CV(%) |
|---|---|---|---|
A | 0.90 | 0.03 | 3.3 |
B | 0.51 | 0.04 | 8.6 |
C | 0.37 | 0.02 | 6.2 |
n=8
Between-analysts reproducibility (brown rice)
Samples | Average (ppb) | SD | CV(%) |
|---|---|---|---|
A | 0.85 | 0.03 | 4.0 |
B | 0.50 | 0.02 | 4.9 |
C | 0.37 | 0.05 | 12.3 |
n=3
Daily reproducibility (brown rice)
Samples | Average (ppb) | SD | CV(%) |
|---|---|---|---|
A | 0.93 | 0.06 | 6.7 |
B | 0.52 | 0.05 | 9.7 |
C | 0.37 | 0.01 | 1.4 |
n=3
Recovery test (brown rice)
Concentration of addition (ppm) | Recovery ratio (%) |
|---|---|
0.030 | 151.0±0.8 |
0.020 | 149.2±14.6 |
0.015 | 138.1±4.5 |
n=3
Results of within-run reproducibility, daily reproducibility, between-analysts reproducibility and recovery tests show that this test kit can quantitatively analyze fenitrothion in produce.
Main kit components / shelf-life
Contents (form) | Volume | Quantity |
|---|---|---|
Antibody coated plate (dry form) | 8 rows x 12 columns | 1 piece |
Standard reagent L (freeze-dried: 0.2ppb) | 1mL (after dissolution) | 2 vials |
Standard reagent H (freeze-dried: 2ppb) | 1mL (after dissolution) | 2 vials |
Enzyme-labeled Fenitrothion reagent (freeze-dried) | 6mL (after dissolution) | 2 vials |
Washing reagent (solution: 10 times concentrated) | 50mL | 1 vial |
Chromogenic reagent (solution) | 13mL | 1 vial |
Stop reagent (solution) | 13mL | 1 vial |
Shelf-life: | 12 months after production | Storage at 2 to 8°C |
|---|