Isoxathion

Overview

Isoxathion is an organophosphorous insecticide and is also used for multiple farm products.  This kit allows an easy, rapid (within 2 hours), sensitive, quantitative. multi-sample analysis of Isoxathion that remains in farm products.  This kit is best applied for inspection at delivery time/prior to processing of primary screening in production sites.

 

 

Features

Comparison with instrumental analysis

While instrumental analysis offers high accuracy, it requires complicated and time- consuming pretreatment techniques, and sophisticated instrument operation as well as large quantities of hexane or acetone as organic solvents.  The simpler pretreatment required for kit saves significant time.  Additionally. only a small quantity of methanol is required as an extraction solvent for analyzing isoxathion. 

          Application examples

The figures show the comparison between standard curves with 10% methanol and reaction curves drawn with methanol extract of isoxathion-added farm products (diluted to 10% methanol equivalent).  The overlapping curves; demonstrate the suitability of this kit for analyzing trace Isoxathion without any interference from the original procuce. 

 


Manufactured by HORIBA

Specifications

The following data represent the basic kit performance. For more detailed specifications, please contact us. 

         Standard curve

 

Cross-reactivity

Pesticides

Cross-reactivity(%)

Isoxathion

100

Isoxathion oxone

4.5

Triazofos

0.3

Piraclofos

0.7

Pyridaphenthionn

< 0.1

Sulphotep

< 0.1

Fenitrothion

< 0.1

Clorotalonil

< 0.1

Iminoctadine albesilate

< 0.1

Captan

< 0.1

Fenpropathrin

< 0.1

Acephate

< 0.1

Buprofezin

< 0.1

Mepanipyrim

< 0.1

Permetorin

< 0.1

Diethofencarb

< 0.1

Within-run reproducibility (tomato)

Samples

Average (ppb)

SD

CV(%)

A

15.2

0.8

5.5

B

7.0

0.6

7.9

n=8

          Between-analysts reproducibility (tomato)

Samples

Average (ppb)

SD

CV(%)

A

17.1

1.9

10.9

B

6.8

0.8

11.8

n=3

          Daily reproducibility (tomato)

Samples

Average (ppb)

SD

CV(%)

A

14.8

1.9

12.8

B

7.9

0.8

9.9

n=3

Recovery test (tomato)

Concentration of addition (ppm)

Recovery ratio (%)

2.9

84.8±21.4

10.0

109.9±16.9

20.0

111.9±5.8

n=3

Results of within-run reproducibility, daily reproducibility, between-analysts reproducibility and recovery test show that this kit can quantitatively analyze isoxiathion in roducte.  

Main kit components / shelf-life 

Contents (form)

Volume

Quantity

Antibody coated plate (dry form)

8 rows x 12 columns

1 piece

Standard reagent L (freeze-dried: 1ppb)

1mL (after dissolution)

2 vials

Standard reagent H (freeze-dried: 20ppb)

1mL (after dissolution)

2 vials

Enzyme-labeled Isoxathion reagent (freeze-dried)

6mL (after dissolution)

2 vials

Washing reagent (solution: 10 times concentrated)

50mL

1 vial

Chromogenic reagent (solution)

13mL

1 vial

Stop reagent (solution)

13mL

1 vial

Shelf-life:

5 months after production