VTM 300
Overview
The HORIBA Jobin Yvon VTM 300 system gives you a unique opportunity to enhance your spectroscopic capability to the Vacuum Ultra Violet region. The system combines the HORIBA Jobin Yvon expertise in the following areas:
Spectroscopy system development
Vacuum design
Toroidal Aberration Corrected Gratings
The VTM offers unique performance in the UV and VUV wavelength range. Equipped with an aberration corrected holographic toroidal grating (1200 g/mm) the VTM 300 allows monochromator or spectrograph operation with the same grating. Thanks to this toroidal grating the VTM 300 presents a unique throughput at 100 nm and below.
Features
Unique Features in UV and VUV Wavelength Range
Toroidal Aberration Corrected Grating
High Throughput
Interchangeable Kinematic Grating Mounts
Focal Length 300 mm
Deviation 70 deg
Spectral Wavelength Range 60-300 nm
Spectral Resolution <4 Angstrom
High Repeatability
Monochromator and/or Spectrograph
Bi-dimensional Detection Adapter
25 mm Flat Field
Removable Entrance and Exit Slit
High Vacuum Compatible (10-6 mbar)
Adjustable or Fixed Entrance and Exit Slits
Aperture: f/7.5
Stepping Motor Drive
Manufactured by HORIBA
Specifications
Spectral Performance
Repeatability test with 253.6 nm Hg line
| Helium Full Range Scan
|
Helium Line 58.4 nm
| Hydrogen Line 121.6 nm
|
Mercury Line 194.2 nm
| Mercury Line 253.6 nm
|