Overview

  • Ideal for clinical research
  • Data management on stand-alone PC.
  • Micro-sampling from whole boold (CBC : 30 µL - DIFF: 53 µL) 
  • Cytochemistry, Impedance (real cell volume measurement) & Optical (analysis of the internal cellular structure by measuring light absorbance) - DHSS* Technology.
  • Perfect homogenization of blood samples with reagents - MDSS** Technology.

Features

Pentra ES 60

  • Throughput: Up to 60 samples/hour
  • Reagents: Only 4 onboard reagents and 1 diluent
  • Perfect differentiation of the 5 WBC sub-populations with DHSS* Technology
  • 3 histograms for RBC, BAS/WBC and PLT together with the 5 DIFF Matrix.
  • Basophils counted through specific channel
  • High resolution matrix includes the determination of 2 additional subpopulations (% and #): Atypical Lymphocytes (ALY***) and Large Immature Cells (LIC***)  

* DHSS : Double Hydrodynamic Sequential System

** MDSS : Multi-Distribution Sampling System

*** RUO parameters (Research Use Only)


Manufactured by HORIBA ABX SAS

Specifications

Technical

Weight

77 lbs / 35 kg

Power supply

100 to 240 V (± 10%) - 50 to 60 Hz.

Dimensions

20.3 x 17.5 x 19 in / 51,5 cm x 44,5 cm x 48 cm

Printer

Laser printer

Throughput

60 tests/hour.

Data management

Colour screen: 15 in. monitor (800 x 600px)

Windows XP Prof.TM 

User-defined flagging limits

Transmit patient & QC to LIS

Uni-directional or bi-directional connection

Communication Protocols: ABX / ASTM

CD-ROM drive / Floppy disk

2 serial ports minimum

USB Keyboard / USB Mouse

Memory

10,000 results + Graphics

Quality Control Management

12 selectable QC files

XB: 60 Operator selectable batches with statistics (20 samples per batch)

Within run

Levey-Jennings Graphs

Logs

Reagents, calibration, maintenance, errors, blank cycle

 

Samples

60 samples/hour (open tube)

Parameters (26)

WBC, NEU# & NEU%, LYM# & LYM%, MON# & MON%, EOS# & EOS%, BAS# & BAS%, ALY# & ALY%*, LIC# & LIC%*, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, PLT, MPV, PCT*, PDW*.

Graphs

High-definition LMNE matrix
PLT, RBC, WBC/BAS curves.

Blood Volume

CBC mode : 30 µL
CBC + DIFF mode : 53 µL

Stability

48 hours post-draw stability

Operating sample mode

Open tube

Identification

Alphanumeric. Bar-code reader (optional)

 * RUO parameters (Research Use Only)

 

Results

Parameters

Presicion (%CV)

Linearity

WBC

< 2.0

0-120 x 10³/µL

RBC

< 2.0

0-8 x 106/µL

HGB

< 1.0

0-24 g/dl

HCT

< 2.0

0-67%

PLT (whole blood)

< 5.0

0-1900 x 10³/µL

PLT (PLT concentrate)

< 5.0

0-2800 x 10³/µL