Application
Pentra ES 60
Overview
- Ideal for clinical research
- Data management on stand-alone PC.
- Micro-sampling from whole boold (CBC : 30 µL - DIFF: 53 µL)
- Cytochemistry, Impedance (real cell volume measurement) & Optical (analysis of the internal cellular structure by measuring light absorbance) - DHSS* Technology.
- Perfect homogenization of blood samples with reagents - MDSS** Technology.
Features
Pentra ES 60
- Throughput: Up to 60 samples/hour
- Reagents: Only 4 onboard reagents and 1 diluent
- Perfect differentiation of the 5 WBC sub-populations with DHSS* Technology
- 3 histograms for RBC, BAS/WBC and PLT together with the 5 DIFF Matrix.
- Basophils counted through specific channel
- High resolution matrix includes the determination of 2 additional subpopulations (% and #): Atypical Lymphocytes (ALY***) and Large Immature Cells (LIC***)
* DHSS : Double Hydrodynamic Sequential System
** MDSS : Multi-Distribution Sampling System
*** RUO parameters (Research Use Only)
Manufactured by HORIBA ABX SAS
Specifications
Technical
Weight | 77 lbs / 35 kg | ||
Power supply | 100 to 240 V (± 10%) - 50 to 60 Hz. | ||
Dimensions | 20.3 x 17.5 x 19 in / 51,5 cm x 44,5 cm x 48 cm | ||
Printer | Laser printer | ||
Throughput | 60 tests/hour. | ||
Data management | Colour screen: 15 in. monitor (800 x 600px) Windows XP Prof.TM User-defined flagging limits Transmit patient & QC to LIS Uni-directional or bi-directional connection Communication Protocols: ABX / ASTM CD-ROM drive / Floppy disk 2 serial ports minimum USB Keyboard / USB Mouse | ||
Memory | 10,000 results + Graphics | ||
Quality Control Management | 12 selectable QC files XB: 60 Operator selectable batches with statistics (20 samples per batch) Within run Levey-Jennings Graphs | ||
Logs | Reagents, calibration, maintenance, errors, blank cycle |
Samples
60 samples/hour (open tube) | |
---|---|
Parameters (26) | WBC, NEU# & NEU%, LYM# & LYM%, MON# & MON%, EOS# & EOS%, BAS# & BAS%, ALY# & ALY%*, LIC# & LIC%*, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, PLT, MPV, PCT*, PDW*. |
Graphs | High-definition LMNE matrix |
Blood Volume | CBC mode : 30 µL |
Stability | 48 hours post-draw stability |
Operating sample mode | Open tube |
Identification | Alphanumeric. Bar-code reader (optional) |
* RUO parameters (Research Use Only)
Results
Parameters | Presicion (%CV) | Linearity |
---|---|---|
WBC | < 2.0 | 0-120 x 10³/µL |
RBC | < 2.0 | 0-8 x 106/µL |
HGB | < 1.0 | 0-24 g/dl |
HCT | < 2.0 | 0-67% |
PLT (whole blood) | < 5.0 | 0-1900 x 10³/µL |
PLT (PLT concentrate) | < 5.0 | 0-2800 x 10³/µL |