Patient Demography: Female aged 72

Symptoms: Bilateral lower limb pain? CML

Diagnosis: Chronic Myeloid Leukemia in Blast Crisis

Microscopy: Blast-70%, Lymphocyte-04%, Monocyte-03%, Neutrophil-06%, Eosinophil-02%, Basophils-05%, Promyelocytes-03%, Myelocytes-04%, Metamyelocytes-03%

Other Tests: BCR ABL t (9:22) Philadelphia Chromosome – POSITIVE, MTHFR Gene mutation – POSITIVE

How Yumizen Can Assist: 1. On the DIFF scattergram, we observe a compact blast population between the lymphocyte, atypical lymphocyte, and monocyte areas. The small to moderate-sized blasts are situated in the ALY area. These blasts have basophilic cytoplasm with irregular nuclear contours and contains 2 to 3 nucleoli. 2. WBC distribution histogram also suggests the presence of basophils, which is confirmed in the manual differential count.

(Picture) Test results monitor screen

Microscopy findings showed small to moderate-size blasts having basophilic cytoplasm and irregular nuclear contours with 2 to 3 nucleoli. Basophils were also seen.

This clinical case has been provided by Dr. Sonal Gupta, Lab Head & Pathologist, Redcliffe Diagnostic Lab, Mumbai, India.

HORIBA Medical Chase the Case #15 - Acute Myeloblastic Leukemia

Subscribe Chase the Case

Subscribe and receive the latest Chase the Case by e-mail every month. Once subscribed, you will start to receive it from the next issue.

Subscribe
Yumizen H2500 Hematology Analyzer-V2 logo-HORIBA Medical

Yumizen H2500/H1500 Hematology Analyzer

Mastering the analyzis process to deliver reliable results

more

Pedido de Informação

Você tem alguma dúvida ou solicitação? Utilize este formulário para entrar em contato com nossos especialistas.

* Esses campos são obrigatórios.

Corporate