光子计数PMT检测系统

独立PMT外壳,用于光谱的定量测量和低光强度成像

在紫外/可见/近红外光谱中,光电倍增管(PMT)是用于低光强度定量测量的理想检测器。PMT灵敏度高,具有非常宽的动态范围,因此还可以测量高光通量的光。PMT的速度也非常快,所以可以监测光信号的快速变化。另外,光电倍增管坚固耐用,使用寿命长。

HORIBA提供简单独立的PMT外壳,通过开关轻松切换模拟或光子计数模式操作,可兼容光谱范围覆盖185 – 900nm的侧向光电倍增管。对于非制冷PMT外壳,它的暗计数非常低,大约每秒50到500个计数(cps)。

事业部: 科学仪器
产品分类: 定制光谱解决方案
制造商: HORIBA Scientific

Key Benefits

  • Compact Size
  • Inexpensive
  • Operates in Analog or Photon Counting Modes
  • Detection from 185 to 900 nm
  • Complete Package
  • LCD display of output signal
  • Easy to Use

PMT's are better than photodiodes and APD's for low light detection

Although the quantum efficiency (QE) of a PMT is less than that of a silicon photodiode or avalanche photodiode (APD), a PMT has a much lower limit of low light detection due to the high current amplification (~ 107) of the electron multiplier in the PMT. It is because of this large gain and low noise that a PMT provides significantly better low light sensitivity than a photodiode or APD.

Looking for NIR Detection?

For NIR detection beyond 900 nm please take a look at our CoolOne cooled PMT housing for photomultiplier detection from 185 to 1,200 nm and also our InGaAs detectors for detection from 0.8 to 1.9 μm.

The HORIBA ambient PMT housing includes the following

  • On/off switch
  • Complete analog and photon counting circuitry
  • Analog or photon counting mode/output switch
  • Line voltage power supply
  • High voltage power supply
  • LCD display of PMT high voltage or measured analog signal
  • Signal out BNC connection
  • High voltage adjustment knob
  • External high voltage control input BNC connection (for optional external control)
  • RF shielded case
  • Internal socket for most 1 1/8-inch side-on photomultiplier tubes

HORIBA's photomultiplier detection system is very convenient and easy to use. It is built as a complete unit using modern design techniques and components. Everything needed to quantitate the intensity of light is packaged into a compact, attractive housing that attaches conveniently to your experiment. There are no unsightly boxes for power supplies or a multitude of cables tangled on your bench-top. The system incorporates a high voltage supply, all detector electronics, and the controls in a compact, attractive housing that also contains the PMT. You have a choice of standard PMTs to choose from or you can select from any other 1 1/8-inch side-on photomultiplier tubes that operate from -200 to -1,100 Volts DC. The PMT high voltage is shown on a digital display on the housing or you can select to show the actual analog signal from the PMT with just a flip of a switch. The high voltage can be adjusting manually or controlled externally with a signal at the “External Voltage Control” BNC. Power is delivered from an external AC adapter that is included.

Controls

All controls and displays are available on the front and side of the OBB PMT housing.

Control Specifications

Display

LCD display of high voltage or signal

Controls

On/Off switch
Analog/Digital Mode switch
Voltage/Signal Display switch
Voltage Adjust potentiometer
Analog Gain selection
Analog Time Constant selection
Analog Offset

BNC Connectors

Signal Out
External Voltage Control

Physical Specifications

Dimensions (WxDxH) 4 x 4 x 8 inches  
Weight 4 lbs  

High Voltage Power Supply Specifications

Input ± 15 VDC, 250 mA  
High Voltage -200 to -1,100 VDC manually adjustable LCD displays actual cathode voltage  
External High 0 to +5 VDC (0 = -200 V, 5 = -1,100 V)  
Voltage Adjust Continuously adjustable  
Input Regulation ± 0.05% max. (for 15 V ± 1 V input).  
Load Regulation ± 0.05% max.  
Ripple 100 mV p-p max.  
Temperature Coefficient ± 0.01% max. (+5 to 40°C)  
Drift ± 0.03%/hr. max. (after 15 minute warm-up)  

Analog Mode Specifications

Gain Settings: 1 µA = 1 V
0.1 µA = 1 V
0.01 µA = 1 V
0.001 µA = 1 V
 
Time Constant Settings:
(with Corresponding Analog Frequency)
0.05 msec (20 KHz)
0.5 msec (2 KHz)
5 msec (200 Hz)
50 msec (20 Hz)
500 msec (2 Hz)
 
Offset Correction: ± 50 nA  
Signal Output on BNC connector: Analog voltage  

Photon Counting Mode Specifications

Linear Dynamic Range: 5 orders of magnitude*  
Dead Time: 250 ns  
Signal Output on BNC connector: TTL  

* Linear dynamic range of the OBB photon counting PMT housing. The intensities were produced by attenuating the fluorescence emission of a fluorescein sample with neutral density filters.

留言咨询

如您有任何疑问,请在此留下详细需求信息,我们将竭诚为您服务。

* 这些字段为必填项。

Related products

CoolOne
CoolOne

双模模拟/光子计数PMT

FHR1000
FHR1000

高分辨率单色仪

Gemini 180
Gemini 180

双级联单色仪

iHR系列
iHR系列

中等焦长成像光谱仪

MicroHR系列
MicroHR系列

短焦长成像光谱仪

Synapse InGaAs/Symphony II InGaAs
Synapse InGaAs/Symphony II InGaAs

深度制冷近红外探测器

光子计数PMT检测系统

独立PMT外壳,用于光谱的定量测量和低光强度成像

单通道探测器
单通道探测器

PMT、固态、光电探测器可选,用于定制光谱解决方案

科研级相机
科研级相机

深度制冷,覆盖紫外/可见/近红外范围