CS-100F1 系列

光纤型化学药液浓度计

CS-100F1 系列在CS-100的基础上配备光纤。样品单元直接集成到于清洗系统的管路,并使用光纤电缆传输光信号,提供实时在线测量。每台浓度计可实现多达四种药液的测量,可以在单槽或单晶圆清洗系统中检测多种化学药液。

可测量化学药液示例

事业部: 半导体
产品分类: 湿法过程控制
制造商: HORIBA Advanced Techno Co., Ltd.
  • 使用在线样品单元实时跟进浓度信息
    使用直接连接清洗系统主管路的在线样品单元实时检测浓度。
  • 多达四种药液的多重检测
    每台浓度计可测量多达四种化学药液或浓度范围。用户可单次检测多种化学药液的浓度(或不同配比的化学药液),如单槽或单晶圆清洗系统。
  • 更高精度的样品单元类型
    为保证稳定性,还提供一种样品单元类型,支持使用浓度计输出信号反馈进行控制。该浓度计还支持将化学药液浓度保持在可接受的范围内,并消除不必要的化学药液变化。
  • 电气单元和测量单元完全分离,安全安装
  • 符合 RoHS 标准
ModelCS-151F1CS-152F1CS-150F1
Target of
measurement
SC-1
(ammonia/hydrogen peroxide
aqueous solution)
SC-2
(hydrochloric acid/hydrogen
peroxide aqueous solution)
SPM
(sulfuric acid/hydrogen
peroxide aqueous solution)
Multi-monitoring with one unit of up to 4 types of chemical solutions or ranges possible.
Measurement principleAbsorption spectrometry
Concentration computation
principle
Temperature-compensated multivariate analysis
Measurement range
(mass %)
NH3: 0.00% to 1.00%
H2O2: 0.00% to 5.00%
H2O: 94.0% to 100.0%
HCI: 0.00% to 2.00%
H2O2: 0.00% to 2.00%
H2O: 96.0% to 100.0%
H2SO4: 70.0% to 96.0%
H2O2: 0.00% to 10.00%
H2O: 4.0% to 30.0%
Repeatability (mass %)
(within ±1% room temperature)
NH3: ±0.15%
H2O2: ±0.30%
H2O: ±1.5%
HCI: ±0.15%
H2O2: ±0.15 %
H2O: ±1.5%
H2SO4: ±0.50%
H2O2: ±0.50%
H2O: ±1.5%
・Repeatability for the list above is for the cell unit type.Please contact us for the repeatability of the in-line cell type.
・Differences in the above list may occur depending on measurement range.(Please see our website for details.)
Sample solution temperature20 to 80: (when using Inline cell) 20 to 30: (when using cell unit)
Measurement cycleApprox. 3 seconds
Content of
LCD and input
• 1st component: measured concentration value (%)   •Error code
• 2nd component: measured concentration value (%)  •Parallel concentration alarm output invalid status
• 3rd component: measured concentration value (%)  •Serial/parallel status
• 1st and 2nd components: concentration alarm status •Background compensation data
RS-232C serial input• Serial/parallel status switching            •Parallel concentration alarm output status
• Measurement data transfer request         (valid/invalid)
• 1st and 2nd component concentration alarm setting  •Target measurement solution switching
Parallel input12 to 30 V DC (photocoupler insulated)        • Parallel concentration alarm output
                          • Target measurement solution switching
                          • Background compensation
Parallel outputOpen collector output (photocoupler insulated)     • 1st component concentration alarm, 4 types
Max. current when on:                 • 2nd component concentration alarm, 4 types
  DC 5 mA (no internal protection resistor)        • Device error
Max. impressed voltage when off: 30 V DC       • Measuring
                           • Target measurement solution type
                           • Background compensation status
Solution temperature inputTerminal block connection (M3 terminal screws)
• Platinum resistance temperature sensor: Pt100 (Pt100 sensor should be prepared by customers)
Analog output• 1st component concentration alarm, 4 types: 4 to 20 mA
• 2nd component concentration alarm, 4 types: 4 to 20 mA
✽We also support other measurement ranges. Please specify when ordering.
Fiber optic cableDedicated CS-100F1 Series fiber optic cable (25 ±15: installation temperature)
Sample cellWe can supply dedicated sample cells (inline cell or cell unit) for various kinds of chemical solutions.
Power supply100 V to 230 V AC (single phase) ±10%, 50/60 Hz, approx. 100 VA
Dimensions205 (W) x 329 (D) x 269 (H) mm 8.1 (W) x 13.0 (D) x 10.6 (H) in (excluding protruding part)
WeightApprox. 11 kg

Note: Please consult us regarding concentration monitors for chemical solutions other than those given above.

Schematics

Concentration monitor

Inline cell                                                                                     Cell unit

              