TB-01
Übersicht
The TB-01 contains a high-speed leading-edge discriminator circuit that can be configured to detect positive or negative pulses and to trigger on rising or falling edges. NIM or TTL-compatible outputs can be selected, and both the threshold voltage and blocking width/output duration can be adjusted.
Applications
The TB-01 pulse converter/discriminator has more uses than we can think of. Here are just a few...
- Use the TB-01 with a photodiode to generate sync pulses from your mode-locked laser with picosecond timing performance
- Convert positive and negative pulses to fast-NIM
- Convert positive and negative pulses to 50Ω TTL
- Convert fast-NIM to TTL and vice versa
- The TB-01 can even be used as a photomultiplier discriminator
Performance Data
Application 1: Laser synchronization
This TCSPC instrument response function from a femtosecond laser was obtained using a photodiode and the TB-01 pulse converter on the Start channel and a MCP-PMT on the Stop channel. The TB-01 was close-coupled to the photodiode on the optical table. The TB-01’s jitter contribution is negligible in this case.
Application 2: TB-01 as photomultiplier discriminator
Comparison of a linear-focused PMT’s instrument response function measured using a TB-01 and a NIM style leading-edge discriminator. FWHMs obtained were 1.60ns for both devices.
Hergestellt von HORIBA Scientific
Spezifikationen
Pulse pair resolving time | 10ns. Maximum counting rate 100MHz. |
Blocking width/output duration | 5ns to 65ns nominal. Internally adjustable. |
Threshold | -1V to +1V nominal. Internally adjustable. |
Outputs | Fast NIM compatible or TTL compatible (2.2V nom peak @ >10ns 50Ω). |
Connectors | NIM standard QLA/Lemo00 for input/output, Lemo 0B303 for power. |
Dimensions | 34mm x 64mm x 58mm. |
Power requirements | +15V @ +200mA. Internally regulated. |