Fluoreszenz Spektroskopie

News

  • HORIBA Scientific Presents Young Fluorescence Investigator Award at Biophysical Show. Read more
  • HORIBA Scientific Announces New Ultima TCSPC lifetime system, the next generation of turnkey TCSPC Fluorescence Lifetime systems...Read more
  • HORIBA Scientific announces the acquisition of the Photon Technology International (PTI) Business... Read more
  • HORIBA Scientific Announces New Generation of Turnkey TCSPC Fluorescence Lifetime Systems...Read more
  • Aqualog CDOM Fluorometer Wins Award as one of Best New Instruments in 2011...Read more
  • HORIBA Scientific and NKT Photonics Introduce the World’s First Commercial Integrated Supercontinuum Powered Spectrofluorometers... Read more
  • Aqualog®, the only simultaneous absorbance and fluorescence system, measures CDOM (colored dissolved organic matter) in water 100 times faster than previous fluorescence methods...Read more
  • World's First 100MHz Picosecond Diode Source With USB Interface and Plug-and-Play Interchangeable Heads ...Read more

Aqualog® - unser kompaktes Tischfluorimeter für CDOM-Analyse

Übersicht

Das AQUALOG ist das weltweit einzige Spektrometer, welches simultan Absorptionsspektren und Fluoreszenz Anregungs-Emissions-Matrices (EEM) misst. Die Aufnahme der EEM ist bis zu 100-mal schneller als mit herkömmlichen Spektrometern. Die nutzerorientierte Software vereinfacht die Messung dank Automatisierung, z.B. mittels Quinin Sulfat Kalibrierung, Korrektur der inneren Filtereffekte sowie der Rayleigh- und Ramanstreuung. Zusätzlich ermöglicht der schnelle und einfache Export der Daten die Anwendung von multivariaten Auswertungen, wie z.B. PARAFAC, zur Identifizierung einzelner organischer Komponenten.

Merkmale

Hardware

  • Das einzig existierende System, das simultan Absorption und Fluoreszenz misst
  • TE gekühlter CCD Fluoreszenz Emissionsdetektor  zur 100 mal schnelleren Datensammlung als herkömmliche Laborfluorimeter
  • Korrigierte UV-VIS Absorptionsaufnahme für Stabilität und Genauigkeit
  • Doppelmonochromator in der Anregung für hervorragende Streulichtunterdrückung
  • Angepasster Bandpass für Absorption und Fluoreszenzspektren
  • Option für automatischen Probenwechsler (2 oder 4 Positionen)
  • Kompatibel mit Durchflusszellen und Titratoreinheit

    Software

    • Optimierte Experiment Menüs um die Konfigurationszeit für Anwender zu minimieren
    • Vollständige und automatische Korrektur der Fluoreszenzspektren mit NIST Standards
    • Spektrale und Kinetische Analysetools für Absorptions- und Fluoreszenzdaten
    • Methoden und Batchprotokolle für automatische mehrfache Probenmessungen

      'How Fast is Fast?'

      Click here for the Aqualog 3.6.11 Patch.
      Please Note: Please select to save the file. Release Notes.


      Hergestellt von HORIBA

      Spezifikationen

      Fluorescence Hardware

      Light source

      150W ozone free vertically mounted xenon arc lamp

      Excitation range

      240-630 nm

      Excitation bandpass

      5 nm

      Excitation monochromator

      Subtractive double monochromator

      Excitation gratings

      1200 gr/mm, 250 nm blaze

      Excitation wavelength accuracy

      ±1 nm

      Emission range

       210-620 nm

      Emission bandpass

      5 nm

      Emission spectrograph

      Fixed, aberration-corrected 140 mm focal length

      Emission grating

      405 gr/mm; 250 nm blaze

      Emission detector

      TE-cooled back-illuminated CCD

      Emission integration time

      5 ms minimum

      Hardware pixel binning

      0.41, 0.82, 1.64, 3.28 nm/pixel

      CCD gain options

      2.25 e-/cts in high gain, 4.5 e-/cts in medium gain, 9 e-/cts in low gain

      Sensitivity

      Water-Raman SNR > 20,000:1 (RMS method) (350 nm excitation, 30s integration)

      Weight

      32.72 kg (72 lbs)

      Dimensions

      LWH (618 x 435 x 336 mm); (24" x 17" x 13")

      Absorbance Hardware

      Scanning range

      230 - 800 nm (optical)

      Bandpass

      5 nm

      Slew speed

      Max 500 nm/s

      Optical system

      Corrected single beam

       Detector

      Si photodiode

      Wavelength accuracy

      ±1 nm

      Wavelength repeatability

      +/- 0.5 nm

      Photometric accuracy

      ±0.01 AU from 0 to 2A

      Photometric stability

      <0.002 AU per h

      Photometric repeatability

       +/- 0.002 AU (0 to 1 AU)

      Stray light

       

      <1% measured with KI standard

      Duetta fluorescence and absorbance