IT-470F-H Built-in type Non-Contact Infrared Thermometer
Overview
The IT-470F-H is a built-in type, non-contact infrared thermometer with industry-leading accuracy. It contributes to improve process stabilization, for which requires high accuracy temperature measurement, such as semiconductor and FPD production.
Features
- Absolute high measurement accuracy ±0.1°C* and superior repeatability(2σ) of 0.1°C*.
*Actual value, under given conditions.
Application
For semiconductor and FPD production which requires high accuracy temperature measurement
- In case of a susceptor is in a vacuum chamber
Temperature measurement of a susceptor is possible from outside of a chamber through the window - In case of surface temperature measurement of liquid crystal substrate in FPD production process.
As non contact measurement, surface is not damaged. - In case of temperature controlling in polishing
As non contact measurement, surface is not damaged.
Manufactured by HORIBA
Specifications
Measurement wavelength | 8 to 14µm |
---|---|
Measurement temp. range | -50 to 200℃ |
Output resolution | Optical digital output: 0.1℃ Current output ≤0.025%FS |
Accuracy* | ±4.0℃ (Object temperature -50℃) |
Repeatability | 0.7℃ (Object temperature -50℃) |
Response time (Optical digital output) | Optical digital output: ≤1.4 sec (running average 1 sec, at 95% response) |
Data sampling period | 0.1 sec |
Display update period | Optical digital output: 0.2 sec Current output: 0.1 sec |
Measuring diameter | Φ8mm/Distance 150mm (incident light amount 90%) |
Emissivity setting | 0.001 to 1.000 changeable (Setting at factory shipping) |
Operating temp. & humidity | 0 to 55℃, 35〜85%RH (no condensation) |
Power | DC24V±5% |
Function | Optical digital output (JIS F05) Current output 4 to 20mA (Load resistance 100Ω, non-isolated) |
Dimensions | 55×44×96mm Power, Current cable attached (Standard length 2m) |
Mass | ≤300g |
*Condition: Operating temperature 24 to 29℃ (include temperature of attachment), Humidity 55±20%, Emissivity correction setting 1.000.