Model | LAQUA EC210 Conductivity/Resistivity/Salinity/Total Dissolved Solids/Temperature (°C/°F) |
---|---|
Conductivity Range | ...µS/cm to 200.0 mS/cm (k=1.0) |
Resolution | 0.05% full scale |
Accuracy | ±0.6% full scale; ±1.5% full scale > 18.0 mS/cm |
Reference Temperature | 15 to 30 °C (adjustable) |
Temperature Coefficient | 0.00 to 10.00% (adjustable) |
Cell Constant | 0.1, 1.0, 10.0 (adjustable) |
Calibration Points | Up to 4 (Auto) / Up to 5 (Manual) |
Measurement Unit | S/cm, S/m (selectable / auto-ranging) |
TDS Range | ...ppm to 100 ppt (TDS factor=0.5) |
Resolution | 0.01 ppm (mg/L) / 0.1 ppt (g/L) |
Accuracy | ±0.1% full scale |
TDS Curves | Linear (0.40 to 1.00), EN27888, 442, NaCl |
Resistivity Range | 0.000 Ω•cm to 20.0 MΩ•cm |
Resolution | 0.05% full scale |
Accuracy | ±0.6% full scale; ±1.5% full scale > 18.0 MΩ•cm |
Salinity Range | 0.0 to 100.0 ppt / 0.00 to 10.00 % |
Resolution | 0.1 ppt / 0.01% |
Accuracy | ±0.2% full scale |
Salinity Curves | NaCl / Seawater |
Calibration Option | Yes |
Temperature Range | -30.0 to 130.0 °C / -22.0 to 266.0 °F |
Resolution | 0.1 °C / °F |
Accuracy | ± 0.5 °C / ± 0.9 °C |
Calibration Option | Yes |
Memory Data | 500 |
Auto Log Data | Yes |
Measurement Modes | Auto Stable, Auto Hold, Real Time |
Auto Shut-Off | Yes (Programmable: up to 30 minutes) |
Electrode Status | Yes (On screen display) |
Diagnostic Messages | Yes |
Meter Inputs | BNC (for conductivity electrode), phono (for temperature sensor) |
Display | Custom LCD with backlight |
Housing | IP67, shock & scratch resistant, non-slip |
Power Requirement | 2 x AA batteries |
Battery Life | > 500 hours |
Meter Dimensions | 160 (L) x 80 (W) x 40.60 (H) mm |
Meter Weight | Approx. 260g (with batteries) / 216g (without batteries) |