PoliSpectra® Quad

Multichannel Grating Spectrometer

Horiba’s OEM Polispectra Quad spectrometer is an integration of four VS70 spectrometers built around Horiba JY Type-IV flat-field aberration-corrected holographic concave gratings. This high throughput VIS spectrometer with 4 fiber inputs provides simultaneous acquisition of 4 channels at high speed with minimal crosstalk. The main electronics board is designed to read the two CCDs on the system simultaneously and provides 1-8X binning capability. This OEM system is a unique and innovative approach to drastically reduce the cost per channel. 

For OEM Industrial Low-light Applications

  • Fluorescence 
  • Emission (OES)
  • Absorbance
  • Reflectance
  • Examples:
    • Body Fluids Analysis 
    • Flow Cytometry
    • LED Testing, Colorimetry
    • Food Analysis Moisture Content

 

Spectroscopy in 60 Seconds - PoliSpectra Quad Video (1:09)

*Available in select countries.

You can email OEM Technical Sales at: OEM.US(at)horiba.com or contact OEM at: +1.732.494.8660 Ext. 7733

Segment: Scientific
Manufacturing Company: HORIBA Scientific

Spectroscopy in 60 Seconds - PoliSpectra Quad Video (1:09)

 

  • Simultaneous acquisition of 4 channels
  • Excellent light purity with a typical 0.05% channel to channel crosstalk
  • Spectral acquisition time as low as 1.3 ms for all 4 channels 
  • High throughput (f/2.3)
  • Ultra-low stray light with aberration correction holographic grating
  • High SNR and dynamic range
  • TTL in/out to trigger each and all channels
  • Compact size, robustness, stability and affordability 

 

CCD Quantum Efficiency

 

Simultaneous measurement of 4 colors LED array spectra with Polispectra Quad spectrometer

Polispectra Quad Spectrometer is capable of simultaneously measuring 4 channels, the picture above shows the emission spectra of 4 different color LEDs measured using Ø 200 µm core fibers
 

Raw Throughput Uniformity

Raw throughput uniformity of better than 75% for 1000 production units measured channel to channel

 

OEM Spectrometer Selection Guide

View and Download our OEM Spectrometer Selection Guide

You can email OEM Technical Sales at: OEM.US(at)horiba.com or contact OEM at: +1.732.494.8660 Ext. 7733

 

General Spectrometer Specifications*

Spectral Coverage

Tunable from 470 to 730 nm

Spectrometer Input

FC (standard), other input types available upon request

Spectral Resolution

2.7 nm with 75 µm slit; 5 nm for  Ø 200 µm fiber (no slit)

Average Spectral Dispersion

23.56 nm/mm; 0.33 nm/pixel (14 µm pixel CCD detector)

Crosstalk (channel to channel)

< 0.05% (typical)+
< 0.1% (maximum)

Focal Length

70 mm

F/#

~ F/2.3

Stray Light Rejection (< 590 nm or > 680 nm)

< 0.05 % (typical) 

Wavelength Accuracy

< 0.25 nm (average between 500 nm and 710 nm)

Software

LabVIEW™ acquisition software for initial evaluation (DLLs provided for software integration)

Back-illuminated linear CCD Detector
CCD Detector Model Hamamatsu S11071-1104N 
CCD Sensor Format 2048 x 22  (2 spectra per CCD of about 1000 pixels each)
CCD Pixel Size  14 x 14 µm
CCD Height  0.3 mm
CCD QE at 20° C QE = 63% at 250 nm; 76% at 650 nm; 55% at 850 nm
Scan Rates 2 MHz
Full Well Capacity > 225,000 e¯/pixel (minimum); 250,000 e¯/pixel (typical)
Communication USB 2
Scan Rates 45 kHz and 500 kHz
Readout Noise 23 e¯ (typical) to 35 e¯ (maximum)
Maximum Spectral Rate 720 spectra/s with 0 ms exposure time 
Digitization 16-bit ADC
Dynamic Range
(typical for single pixel)
11,000:1 (typical)
7,300:1 (minimum)
Non-linearity
(measured on each system)
<0.8% (typical)
1.5% (maximum)
Dark Current at –45° C                                 
Note: Pixel Size = 14 μm
50 e¯/pixel/s (typical)
Environmental Conditions Operating temperature 15° C to 35° C ambient
Relative humidity <70% (non-condensing)
Storage temperature -25° C to 45° C
Power requirements
AC/DC power supply (provided)
90-264 VAC, 47–63 Hz

*Specifications, form factor, and spectrometer cover subject to change without notice. 
No LabVIEW™ license is needed to run our acquisition software.

+ This value includes optical and electronic crosstalk

 

Mechanical Drawings

Download Mechanical Drawings

 

You can email OEM Technical Sales at: OEM.US(at)horiba.com or contact OEM at: +1.732.494.8660 Ext. 7733

Request for Information

Do you have any questions or requests? Use this form to contact our specialists.

* These fields are mandatory.