ABX Pentra DX 120 SPS Evolution

In the heart of hematopoiesis

Product Description

  • Autoloader with continuous loading capability
  • Double DIFF Matrix
  • Automatic reflex testing
  • Automatic validation of results
  • DHSS* technology (*HABX patent)

ABX Pentra DX 120

  • Methods of measurement: Flow cytometry, Cytochemistry, impedance, absorbance, optical cytometry and fluoro-flowcytometry
  • QUADIMENSION Concept: Automatic correction of the leukocyte count based on 4 independent principles
  • Automatic reflex testing: Selective and programmable (hematology parameters alarms and flags)
  • Rack rotation mixing of samples: Smooth and efficient
  • Liquid valves (no pinch valves): Reliability and increased precision

SPS Evolution

  • Drop needle system: Auto-cleaning drop needle system. Calibration of blood drop.
  • Smearing system: No inter-sampling contamination, No carry-over, Easy to replace.
  • Programmable staining protocols: MGG, Wright, Wright Giemsa
Segment: Medical
Division: Hematology
Base product
Manufacturing Company: HORIBA ABX SAS

49 parameters

Integrated slidemaker-stainer

Comprehensive validation system

 

Technical

Weight374 lb
Power supply

100/240 V - 50/60 Hz - 700W
laser in stand-by 800 W, (2kW maxi).

Dimensions29.5 in x 65 in x 22.6 in (H x W x D)
PrinterLaser
Throughout120 samples/hour
Data processing 
  • Windows XP, Floppy disk,
  • CDROM (R/W), RS 232C, TCP/IP,
    2 x USB1 Mono and hi-directional connection
 
ComplianceASTM
Memory capacity80 GB unlimited patient database
Patient ManagementAutomatic validation (setting rules), Data Check, Patient history (matrix, curves, data), manual entry.

 

Samples

Parameters (49) WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, PLT, MPV, *PCT, *PDW %, NEU, LYM, MON, EOS, BAS, *ALY, *LIC, *IMM, *IML, *IMG : (% and #), RET (% and #), RETH %, RETM %, RETL %, *IMM %, CRC%, IRF %, MRV, MFI. NRBC % NRBC#, *CWBC, *WBC#, *RBC#, *PolyNuc#, *MonoNuc#
*RUO (Research Use Only)
Volume

Manual mode : 130 µL
Automatic mode : 200 µL 

Slide50 uL automatic cycle
StabilityUp to 48 hours for CBC/DIFF
Operating sample modeAutoloader 15 racks (150 tubes) capacity
Continuous feed for 8-racks, 160 slides capacity
Sample mixing with end-over-end rack rotation
Open tube in STAT mode
IdentificationAlphanumeric, Bar-code reader (internal and external)
Automatic 3-line printing on the slide using indelible ink
Flags : 46 possible pathologic flags

 

Results

ParametersPresicion (cv %)Linearity
WBC< 2%0-150 x 103 / mm3
RBC< 2%0-8 x 106 / mm³
HGB< 1%0-24 g/dL
HCT< 2%0-67%
PLT (whole blood)< 5%0-1900 x 10³ / mm³
PLT (PLT concentrate)< 5%0-2800 x 10³ / mm³
LYM< 5%--
MON< 15%--
NEU< 3%--
EOS< 25%--
BAS< 40%--

 

Blood Control

Reference

Description

 

2062207

ABX Difftrol Twin pack - 2L 

2 x 3 mL vial

2062203

ABX Difftrol Twin pack - 2N

2 x 3 mL vial

2062208

ABX Difftrol Twin pack - 2H

2 x 3 mL vial

2072201

ABX Minotrol Retic Twin pack - 2N

2 x 3 mL vial

2072202

ABX Minotrol Retic Twin pack - 1L  1H

2 x 3 mL vial

2072203

ABX Erytrol Twin pack 2 x level 3

2 x 3 mL vial

2072204

ABX Erytrol Twin pack 1 x levels 1 & 2

2 x 3 mL vial

Calibrators

Reference

Description

 

2032002

ABX Minocal

1x2 mL

N: normal level, H: high level, L: low level
* Recommended by HORIBA Medical

REQUEST FOR INFORMATION

Do you have any questions or requests? Use this form to contact our specialists.

Product Accessories

Hematology Reagents_FR
MoreABX Pentra DX 120 SPS Evolution Reagents
MoreABX Pentra ML
Centralized Validation
MoreEasyCell
EasyCell® assistant - Cell Image Analysis for the Hematology Laboratory
MoreHemaprep
Automated Blood smearing