Ambient Hydrogen Sulfide Monitor APSA-370/CU-1

APSA-370/CU-1 Ambient H2S monitor measures SO2 converted H2S through oxidation catalyst based on Ultra Violet Fluorescence. In order to reduce measurement error by water concentration fluctuatin in ambient, humidifier is set at sample line. It enables long-term stability according to stabilizing catalytic reaction.

Segment: Process And Environment
Division: Ambient
Base product
Manufacturing Company: HORIBA Ltd.
  • Monitor H2S with high sensitivity range in ambient air continuously.
  • Set humidifier at sample line to satbilize catalytic reaction.

 

 

Measurement component 

Hydrogen sulfide (H2S) in ambient air
Measurement principleOxidation catalyst + Ultra Violet Fluorescence
Measurement range0-0.1/0.2/0.5/1.0 ppm
Lower detectable limit2 ppb
Repeatability±3.0% of F.S.
Linearity±2.0% of F.S.
Zero drift±2% of F.S./day
Span drift±5% of F.S./day
Response time (T90)180 sec. or less (from CU-1converter unit)
Sample gas flow rateapprox. 0.7L/min
Operation gasNone
IndicationMeasuring value, alarm
AlarmAPSA-370: Calibration error, Battery error, Flow rate error, Pressure error, Temperature error in catalyzer, etc.
CU-1: Temperature error in catalyzer
Inpt/Output0-1V/0-10V/4-20mA (2 systems: either (1) momentary value and integrated of (2) moving average value)
Contact input/output
RS-232C (option)
Temperature/Humidity0-40°C / 85% or less
Power100/110/115/120/220/230/240 VAC, 50/60Hz
Power consumptionAPSA-370: 150VA
CU-1: 150VA
DimensionAPSA-370: 430(W) x 550(D) x 221(H) mm
CU-1: 430(W) x 550(D) x 221(H) mm
MassAPSA-370: approx. 19kg
CU-1: approx. 10kg