Field-installation Type Electric Resistivity Meter (Four-Wire Transmission)
HE-200R connects resistivity sensor (ERF Series) and measures resistivity and temperature in the sample water. It is suitable for measurement of UPW resistivity.
Resistivity: 0.00 - 20.00 MΩ・cm, 0.00 - 200.0 kΩ・m
Temperature: 0.00 - 100.00 ℃
Product name | Resistivity meter | |||
---|---|---|---|---|
Model | HE-200R | |||
Combined sensor | 2-electrode method resistivity sensor (ERF-001 Series) of cell constant 0.01/cm | |||
Measuring range | Resistivity | MΩ•cm | 0.000 to 2.000 | 0.00 to 20.00 (*Note 1) |
kΩ•cm | 0.00 to 20.00 | 0.0 to 200.0 (*Note 1) | ||
Temperature | ℃ | 0℃ to 100℃ (Display range: −10℃ to 110℃) | ||
Display resolution | Resistivity | As shown in the measuring table. | ||
Temperature | 0.01℃ | |||
Performance | Resistivity | Repeatability | Within ±0.1% of the full scale (response for equivalent input) | |
Linearity | Within ±0.5% of the full scale (response for equivalent input) | |||
Temperature | Repeatability | ±0.1℃ (response for equivalent input) | ||
Linearity | ±0.5℃ (response for equivalent input) | |||
Transmission output | Number of output | 2 (The negative terminals for transmission outputs are internally connected at the same electric potential) | ||
Output type | 4 mA to 20 mA DC: input/output isolated type | |||
Load resistance | Maximum: 900Ω | |||
Linearity | Within ±0.08 mA (output only) | |||
Repeatability | Within ±0.02 mA (output only) | |||
Output range | Output 1 | Resistivity: Free setting within a measuring range. | ||
Output 2 | Temperature: Free setting within a range between −20 and 130℃ | |||
Occasional out for error | Hold or burnout to either 3.8 mA or 21 mA | |||
Transmission hold | In the maintenance mode, transmission signal is held at the latest value or preset value. In the calibration mode, transmission signal can be alive or held. | |||
Contact output | Number of output | 3 | ||
Output type | No-voltage contact output | |||
Contact type | Relay contact; SPDT (1c) | |||
Contact capacity | 250 V AC, 3 A; 30 V DC, 3 A (resistance load) | |||
Contact function | R1, R2 | Selectable from upper limit alarm, lower limit alarm, and currently holding of transmission output. (The contact is closed during alarm operation, opened normally and while the power is down.) | ||
FAIL | Error alarm (Closed in the normal state, opened in the failure state or while the power is down.) | |||
Alarm setting range |
| |||
Contact input | Number of input points | 1 | ||
Contact type | No-voltage "a" contact for open collector | |||
Conditions | ON resistance: 100 Ω max. Open voltage: 24 V DC Short-circuit current: 12 mA DC max. | |||
Contact function | External input for transmission holding. | |||
Transmission capability | Method | RS-485 | ||
Signal type | 2 wire system, isolated from the input circuit Not isolated from transmission circuit | |||
Temperature compensation | Applicable temperature element | Platinum resistor: 1 kΩ (0℃) | ||
Compensation method |
| |||
Temperature compensation range | 0℃ to 100℃ (However, the compensation calculation is extended lower than 0℃ or higher than 100℃.) | |||
Calibration | Resistivity | Based on parameter input of coefficient for the sensor cell constant. | ||
Temperature | Comparison to the accurate thermometer Both deviation and coefficient of RTD are taken into account for calibration. | |||
Other function | UPW standard resistivity input | Select one from the next. 182.3 kΩ•m (standard), 181.8 kΩ•m, 182.4 kΩ•m, 18.23 MΩ•cm (standard), 18.18 MΩ•cm, or 18.24 MΩ•cm | ||
Clipping function | If the measured value exceeds the user setting value for clipping, the measured data will be held at the clipping value. | |||
Self-check | Sensor diagnostic error | Temperature sensor short-circuit, temperature sensor disconnection, and out of the temperature measurement range | ||
Converter error | CPU, ADC, and memory errors | |||
Operating temperature range | −20℃ to 55℃ (without freeze) | |||
Operating humidity range | Relative humidity: 5% to 90% (without condensation) | |||
Storage temperature | −25℃ to 65℃ | |||
Power supply | Rated power supply voltage | 100 V to 240 V AC ±10% 50/60 Hz | ||
Power consumption | 15 VA (max) | |||
Other | With power switch for maintenance use | |||
Compatible standards | CE marking | EMC :EN61326-1 Class A, Industrial electromagnetic environment Safety :EN61010-1 RoHS :EN50581 9. Industrial monitoring and control instruments | ||
FCC rules | Part15 Class A | |||
Structure | Installation | Outdoor installation type | ||
Installation method | Mount on 50 A pole or wall | |||
Protection code | IP65 | |||
Case material | Aluminum alloy (coated with epoxy-denatured melamine resin) | |||
Material of fittings | SUS304 | |||
Material of hood | SUS304 stainless steel (coated with epoxy denatured melamine resin) | |||
Material of window | Polycarbonate | |||
Display element | Reflective monochrome LCD | |||
External dimensions | 180 (W) mm × 155 (H) mm × 115 (D) mm (excluding cover and mounting brackets) | |||
Mass | Main body: Approx. 3.5 kg; cover and mounting brackets: Approx. 1 kg |
Do you have any questions or requests? Use this form to contact our specialists. * These fields are mandatory.
Carbon Sensor Conductivity Meter (Low concentration type)
Industrial resistivity meter
KOH Monitor
Wide range TMAH Concentration Monitor
High Precision TMAH Concentration Monitor
Flat Carbon Sensor Conductivity meter
Carbon Sensor Resistivity Meter
Carbon Sensor Conductivity Meter (Low concentration type)
Industrial resistivity meter
KOH Monitor
Wide range TMAH Concentration Monitor
High Precision TMAH Concentration Monitor
Flat Carbon Sensor Conductivity meter
Carbon Sensor Resistivity Meter
2-Channel Resistivity Meter