Oil Content Analyzer OCMA-500

Fully automated one-touch operation, from oil extraction to sample measurement and draining

The OCMA-500 Oil Content Analyzer has been redesigned for even better operability, maintaining its user-friendly features. After injecting the sample, all you have to do is press a button to get the monitoring operation done quickly, from oil extraction to sample measurement and draining. This machine is easy for anyone to use.The OCMA-500 cuts solvent consumption by 20% compared with our previous products, reducing environmental impact. It also reduces the running costs.

 

 

Segment: Process And Environment
Division: Water Pollution
Base product
Manufacturing Company: HORIBA Advanced Techno Co. Ltd.

How to measure:

OCMA-500 Oil Content Analyzer - Measurement

 

  • Fully automated one-touch operation, from oil extraction to sample measurement and draining

  • Reduced environmental impact and running costs
    (20% cut in solvent consumption)

  •  With no need for a solvent evaporation process, the OCMA-500 offers easy operation, and can detect even oils with low boiling points.

 

Improvement of operability:

  • Backlit extraction tank (patent applied for)

The extraction tank is equipped with LEDs. Illuminating the tank makes it easy to check the phase separation between sample and solvent.

  →   

Before extraction                 After extraction
*The color depends on the sample.

  • Color graphic LCD

The OCMA-500 comes equipped with a 3.5-in color graphic LCD. Menus and measurement data can be clearly seen

  • USB flash memory drive port

Storing data on USB flash memories enables easy data management on PCs.

  • Unit conversion

Inputting measurement conditions can change the units displayed (mg/L, mg/kg, mg/g, and mg/PC) as the user needs.

 

Various Application:

  • Factory dranage

For monitoring final discharge water

Factory dranage

  • Ballast water, Bilge

For monitorng discharge from ships

Ballast water, Bilge

  • Vehicle mentenance site and gas station

For monitoring water quality in sorrounding areas

Vehicle mentenance site and gas station

  • Others

Water quality survay based on environmental standard.
For monitoring final discharge from petroleum refinery plant.
For oil dispersion reserch at time of an accident.

Others

 

 

Measurement principle:Solvent extraction, NDIR
analysis (Infrared spectroscopy)
Measurement range:0mg/l ~ 200mg/l dynamic ranging
Resolution0 to 99.9 mg/L; 0.1 mg/L
100 to 200 mg/L; 1 mg/L
Repeatability0 to 9.9 mg/L; ± 0.2 mg/L ± 1 digit
10.0 to 99.9 mg/L; ± 2 mg/L ± 1 digit
100 to 200 mg/L; ± 4 mg/L ± 1 digit
*For standard liquids 
Display Method3.5 inches, 320×240 dots
Backlight Colour graphic LCD
Calibration MethodZero, Span Calibration
Extraction SolventS-316 solvent
Amount of test sample requiredSample water : Solvent = 2 : 1 
Amount of extraction solvent required8 mL
Extraction MethodBuilt-in extractor
Ambient Operationg Temperature0 to 40 ℃ (non condensation)
Power supplyAC 100V to 240V ±10%, 50/60 Hz
Power consumptionAC 100V:Approx.60VA
AC 240V: Approx.90VA
External Dimensions200(W)×313(D)×342(H) mm
MassApprox. 7 kg
External OutputOutput to an USB memory stick
Functions

300-item data memory
Stabilized measurement value display
Self error determination
Clock

Schematics

Brochure-Oil Content Analyzer OCMA-500_550
CategoryDocuments
Size 1.2 MB
FiletypePDF

REQUEST FOR INFORMATION